  IE-LOG-S0842-01 

Instruction Manual for use of S0842 logger  

 

Instrument  is  designed  for  measurement  and  record  of  temperature  from  three  external 

temperature  probes  and  one  binary  signal  from  external  contact.  Temperature  probes  and  binary 
signal  are  connected  by  means  of  connectors.  Measured  value  of  temperature  and  actual  state  of 
binary  input  are  displayed  on  dual  line  LCD  display  and  are  stored  in  adjusted  time  interval  to 
internal  non-volatile  memory.  All  logger  settings  and  control  are  performed by means  of personal 
computer and password protection is enabled. It is also possible to switch ON and switch OFF the 
logger by start/stop magnet (this function can be disabled from the PC). Also automatic switch ON 
at programmed day and time (maximum one month forward) is enabled. Start/stop magnet enables 
also to clear the minimum and maximum value memory. Switching ON a switching OFF the logger 
is possible to control by external binary signal of the 4th channel. 

Minimum  and  maximum  measured  values  can  be  displayed  (display  switches  to  actual 

measured  values  and  min/max  values  automatically).  It  is  also  possible  to  operate  logger  with 
switched OFF display. Short display of actual measured values is enabled by means of magnet. 

Switched ON logger every 10 seconds (independently on logging interval) updates MIN/MAX 

memory,  compares  measured  values  of  each  channel  including  temperature  difference  with  two 
adjustable  limits  for  each  channel  and  exceeding  of  limits  is  indicated  on  the  display  (alarm 
function).  Also  memory  alarm  mode  is  selectable,  when  alarm  is  indicated  permanently  till  alarm 
memory reset. Alarm function is enabled or disabled for each channel individually. Binary channel 
has no alarm mode. 

Logging  mode  can  be  adjusted  as  non-cyclic, when logging stops after filling the memory. In 

cyclic mode oldest stored values are overwritten by new. In addition logging mode can be selected 
when logging is active only if measured value is out of adjusted alarm limits. 

Stored  values  can  be  transferred  from  logger  memory  to  the  PC  by  means  of  communication 

adapter. Communication adapter can be connected to the logger permanently  – data logging is not 
interrupted even if data download appears.  

Logger evaluates minimum battery voltage and its drop below allowed limit is indicated on the 

display.  At  the  same  time  value  of  remaining  battery  capacity  is  available  by  means  of  the  PC 
program and appears on the logger LCD in % (every time after switch ON). 

  

Technical parameters: 

 

 
External temperature probes (can be limited by concrete applied probe):  

Range of measurement:  -90 to +260 °C (RTD Pt1000/3850ppm probe) 

   

-50 to +150 °C (RTD Ni1000/6180ppm probe) 

   

Resolution: 0.1 °C 

   

Accuracy temperature inputs without probe (Pt1000 and Ni1000): 

±0.2 °C from -50 to +100 °C 
±0.2 % from reading from +100 to +260 °C 
±0.4 % from reading from -50 to -90 °C 

Accuracy of the instrument with connected external probe is defined by above input 
accuracy and the accuracy of connected probe. 

Type of compatible probe: based on RTD Pt1000/3850ppm or Ni1000/6180ppm equipped  
  with ELKA3008 female connector 
Recommended probe Pt1000TG8/E (IP67, -80 to +200°C) 
Response time of  xxxTG8/E probe (air flow approximately 1 m/s): 
  t63 < 1 min, t90 < 3 min (temperature step 20 °C) 
Signal for binary input: from potential-less contact, connected to connector ELKA 3008V 
Minimum pulse duration at binary input: 500 ms (shorter pulses may not be recorded) 
Maximum frequency at binary input: 0.5 Hz (i.e. maximum 5 pulses in 10 s) 

Summary of Contents for S0842

Page 1: ...LOGGER S0842 Three channel temperature logger with one binary input and LCD display Instruction Manual...

Page 2: ...tion is enabled or disabled for each channel individually Binary channel has no alarm mode Logging mode can be adjusted as non cyclic when logging stops after filling the memory In cyclic mode oldest...

Page 3: ...of internal RTC 200 ppm i e 0 02 17 28 s in 24 h Power Lithium battery 3 6 V size AA Battery life typical data download to PC is carried out about once a week 5 years in continuous on line mode with...

Page 4: ...eck or optionally set the logger real time clock select suitable logging interval it has no influence to binary inputs record select logging mode cyclic or non cyclic set for each channel type of temp...

Page 5: ...x memory could be stored Notice If logger operates as permanently connected to the computer using of magnet start stop is disabled To enable the logger control by magnet is suitable only in cases when...

Page 6: ...icates logger is ON If information on actual measured values is needed it is possible anytime to display reading display by means of the magnet only if communication adapter is not connected permanent...

Page 7: ...ndicated by INIT reading It can be displayed for about 12 s If battery voltage drop occurred since last internal clock setting below critical limit or battery disconnection for longer period than appr...

Page 8: ...ot possible to combine this function with function of switching logger OFF by magnet User software enables to select only one of them or none Plug magnet to guide slots from logger front side and wait...

Page 9: ...for temperature probe and binary contact It is possible to order as an optional accessory connector for connection of temperature probe Signal cable should be shielded round cross section of minimum d...

Reviews: