  IE-LOG-S0842-01 

Switched  ON  logger  permanently  (with  10  s  interval)  updates  memory  of 
minimum  and  maximum  values  of  temperature  channels  (not  relevant  for 
binary  inputs).  If  displaying  of  MIN/MAX  values  is  selected,  minimum 
measured  value  is  displayed  at  the  LCD  lower  line  (indicated  by  symbol 
MIN)  and  then  similarly  maximum  measured  value  (indicated  by  symbol 
MAX).  In  displaying  of  channel  4  state  always  actual  binary  states  are 
displayed  from  input  4.  All  display  cycle  repeats  periodically,  i.e.  actual 
temperature readings follow. 
 

If  display  is  switched  OFF,

  all  above  readings  are  displayed  up  to 

estimated remaining battery capacity and then display goes out. If logger is 
switched  ON  symbol  LOG  is  displayed  (it  blinks  if  memory  occupation  is 
higher than 90%). 
 
If display is OFF and logger is in mode when record runs only when alarm is 
active,  the  LOG  symbol  is  replaced  by  adjacent  symbol  "–"  (hyphen).  It 
appears in case, all measured values are inside of adjusted alarm limits and 
data  logging  therefore  does  not  run.  Displayed  symbol  indicates  logger  is 
ON. 

 
If  information  on  actual  measured  values  is  needed,  it  is  possible  anytime  to  display  reading 

display by means of the magnet (only if communication adapter is not connected permanently). Plug 
magnet  into  guide  slots  from  logger  front  side  for  approximately  4  s  and  wait  till  reading  on  the 
display appears. If logger has enabled the function switch OFF by magnet, resp. MIN/MAX memory 
clear  by  magnet,  do  not  remove  magnet  from  guide  slots  before  decimal  point  symbol  goes  out  – 
logger  would  be  switched  OFF,  resp.  MIN/MAX  memory  would  be  cleared!  Display  reading 
initialized  by  magnet  goes  out  automatically  after  30  s.  Remove  the  magnet  from  slots  anytime 
during actual reading is ON or later. 

 
 

Temporary displaying of actual reading by magnet 
 

 

 
 

 

Alarm indication on the display 

It is necessary to enable alarm function from the PC and set for each quantity lower and upper 

limit  (not  relevant  for  binary  input).  If  measured  value  is  inside  of  set  limits,  alarm  of  proper 
quantity  is  not  active.  If  value  of  measured  quantity  gets  outside  of  set  limits,  alarm  of  proper 
quantity  is  active  and  it  is  indicated  on  the  display.  It  is  possible  to  select  "memory  alarm  mode" 
when alarm is indicated permanently up to reset from the PC. 

 

Summary of Contents for S0842

Page 1: ...LOGGER S0842 Three channel temperature logger with one binary input and LCD display Instruction Manual...

Page 2: ...tion is enabled or disabled for each channel individually Binary channel has no alarm mode Logging mode can be adjusted as non cyclic when logging stops after filling the memory In cyclic mode oldest...

Page 3: ...of internal RTC 200 ppm i e 0 02 17 28 s in 24 h Power Lithium battery 3 6 V size AA Battery life typical data download to PC is carried out about once a week 5 years in continuous on line mode with...

Page 4: ...eck or optionally set the logger real time clock select suitable logging interval it has no influence to binary inputs record select logging mode cyclic or non cyclic set for each channel type of temp...

Page 5: ...x memory could be stored Notice If logger operates as permanently connected to the computer using of magnet start stop is disabled To enable the logger control by magnet is suitable only in cases when...

Page 6: ...icates logger is ON If information on actual measured values is needed it is possible anytime to display reading display by means of the magnet only if communication adapter is not connected permanent...

Page 7: ...ndicated by INIT reading It can be displayed for about 12 s If battery voltage drop occurred since last internal clock setting below critical limit or battery disconnection for longer period than appr...

Page 8: ...ot possible to combine this function with function of switching logger OFF by magnet User software enables to select only one of them or none Plug magnet to guide slots from logger front side and wait...

Page 9: ...for temperature probe and binary contact It is possible to order as an optional accessory connector for connection of temperature probe Signal cable should be shielded round cross section of minimum d...