background image

Logger operation 

Logger comes complete with installed battery and switched OFF. Before operation it is 

necessary by means of installed user PC software to set logging parameters and other features. The 
PC program is common for all types of loggers Sxxxx – 

for type S7021 it is necessary to have 

installed version 1.10.9.0

 or higher. Older program version can cause logger reconfiguration 

together with loss of proper function! Installed program version can be verified in menu Help

 / 

About program…

 Free newest program version is available to download at www.cometsystem.cz . 

For communication with the PC a communication adapter is necessary (not included in delivery). 
For connection via RS232 serial port it is necessary to use COM ADAPTER, for connection via 
USB port it is necessary to use USB ADAPTER. Connect adapter connector to proper computer 
port and plug the adapter to the guide slots on the side of logger.  
 

 Connecting logger to the computer by means of communication adapter 

 

Notice:

 USB connector can be located also at the computer front side 

 
After connecting the logger to the computer reading of logger info is enabled by means of the 

PC software and also setting of the instrument accordingly to the user needs (menu 

Configuration / 

Setting of instrument parameters

). Before logging start it is necessary: 

 

 

check or optionally set the logger real time clock 

 

select suitable logging interval (it has no influence to binary inputs record) 

 

select logging mode (cyclic or non cyclic) 

 

set counter mode (16 or 32 bits) 

 

specify recalculation constants for counter (if counter reading should indicate directly the 
monitored physical value) and number of digits right from decimal point 

 

switch ON the binary input, if it should be recorded 

 

switch ON the logger (or switch OFF, if delayed start by magnet or automatic is required) 

 

enable or disable the option to switch ON the logger by magnet 

 

enable or disable the option to switch OFF the logger by magnet 

 

set date and time of logger automatic switching ON logger or disable this option 

 

optionally enable to control logging by external signal (i.e. logger switch ON/OFF is 
controlled by binary input state at channel 4). Set required binary state for switching ON the 
logger from the menu of channel 4 parameters. 

 

select if record runs permanently or only if alarm is active 

 

If alarms are about to be applied, set both limits and enable alarm (only for counter input) 

 

optionally enable permanent alarm indication (alarm with memory) 

 

switch ON or switch OFF the display of logger 

 

  IE-LOG-S7021-04 

Summary of Contents for S7021

Page 1: ...LOGGER S7021 Dual channel logger with counter and binary inputs Instruction for use...

Page 2: ...aximum value and logger is switched OFF with the message on the LCD If needed counter mode can be selected when after fulfilling counter overflows and continues from value 0 Logging mode can be adjust...

Page 3: ...me clock adjustable from computer integrated calendar including leap years Error of internal RTC 200 ppm i e 0 02 17 28 s in 24 h Power Lithium battery 3 6 V size AA Battery life typical data download...

Page 4: ...setting of the instrument accordingly to the user needs menu Configuration Setting of instrument parameters Before logging start it is necessary check or optionally set the logger real time clock sele...

Page 5: ...e logger LCD indication ON and OFF is displayed permanently Selection of display of increments at the card of counter channel setting causes record of counter increments in time specified by logging i...

Page 6: ...mperature range battery life can be reduced to 75 i e if displayed remaining battery capacity drops below 25 it is recommended to replace the battery If LCD is switched ON actual readings are displaye...

Page 7: ...ation decimal point goes out logger would be switched OFF Display goes out automatically after 30 s Remove magnet from slots anytime during display is ON or later Activation of switched OFF LCD for 30...

Page 8: ...clock setting below critical limit or battery disconnection for longer period than approximately 30 s after display switch ON during date and time display all four arrows appears as warning to check o...

Page 9: ...s identical with the above procedure for switching ON If decimal point does not appear after 1 s it is necessary remove the magnet and repeat the procedure Logger connection to monitored device by the...

Page 10: ...l logger settings remain unchanged In the opposite case check by means of PC program all settings especially the real time clock in logger Attention inserted battery with incorrect polarity causes log...

Reviews: