22
IE-LGR-Uxxxx-16
up to 10 minutes (option available in the COMET Vision
SW).
We recommend calibrating the relative humidity
and temperature of this device in the laboratory with
a sufficient air flow (at least 1 m / s), the reading can be
performed only after thorough settling, which can take up
to 4 hours.
The device has not a protection against the ingress of
water or dust. Do not use device at places where such
conditions can be expected.
U3631
Compact thermometer -
hygrometer with optional external
temperature probe
This model is designed to measure temperature, relative
humidity and dew-point temperature by internal sensor,
moreover one additional external temperature probe
Pt1000/E can be connected. Internal sensors are
located under the grid on the front of the device.
The device
also
measures
the
temperature
from an external probe and the difference between
this temperature and the dew point temperature.
If the surface temperature of the material is measured
by an external probe, the risk of condensation of water
vapor on the surface to be measured can be determined
immediately.
The device alone is suitable for measuring
in places where there is no rapid change in temperature
or relative humidity, and there is no condensation
of water vapours. If water vapour condensation occurs
inside the datalogger, then the resulting water will
remain there and may damage its electronics.
The device is placed directly in the measured area. The
battery in the device cannot be charged.
We recommend calibrating the relative humidity
and temperature of this device in the laboratory with
a sufficient air flow (at least 1 m / s), the reading can be
performed only after thorough settling, which can take up
to 4 hours.
The maximum length of each probe lead should not
exceed 15 m. It is recommended to use shielded cables.
To maintain the IP rating, the unused probe connector
should be provided with a supplied closing cap.
Summary of Contents for U0110
Page 5: ...IE LGR Uxxxx 16 5 General view of a Uxxxx series datalogger U0141 model...
Page 64: ...64 IE LGR Uxxxx 16 Dimensions U0110 U0111 U0122...
Page 65: ...IE LGR Uxxxx 16 65 U0121 U0141...
Page 66: ...66 IE LGR Uxxxx 16 U0246 U0141T U0541 U5841 U6841 U7844 U2422...
Page 67: ...IE LGR Uxxxx 16 67 U3121 U3120 U4130...
Page 68: ...68 IE LGR Uxxxx 16 U3430 U4440 U3631...
Page 69: ...IE LGR Uxxxx 16 69 U8410...