Before putting the device into operation, it should be considered, if its
usage is suitable for the intended purpose. In connection with this
consideration the device‘s optimum settings should be determined.
In the case that the device is part of a larger measurement system,
instructions for its metrological and operational checks should be
developed
Inadvisable or hazardous applications:
This datalogger is not intended for such applications in which the failure
of its functioning could directly endanger life and health of humans and
animals, or the functioning of other equipment with life-sustaining
functions. In applications the failure or malfunction of which could result
in severe property damage it is recommended to provide the system
with suitable and independent signalling equipment which is able to
evaluate such a condition and, in case of failure, to prevent the above
mentioned damages (See the chapter „
Safety measures and
unauthorized
manipulations“
).
Device positioning:
Adhere to principles and procedures mentioned in this manual. Try to
choose such a place to position the device, where the negative
influence caused by environment is as low as possible. When
performing measurements in refrigerators, metal boxes, metal
chambers and the like, it is every time advisable, as far as the devi
ce’s
operation and reliability is concerned, to place the device outside,
leaving only the sensors and probes inside the measured environment.
Positioning of temperature sensors:
These sensors should be placed in locations where sufficient air
circulation is ensured and where the most critical location is supposed
(according to the application requirements). To prevent heat
conduction through the sensor cables from undesirable influencing
the measurement value, the sensor must be properly inserted into
the measured environment.
If you follow the temperature distribution in
an air-conditioned storehouse, do not place the sensor into the direct
air stream generated by the air-conditioning unit. As a matter of fact,
the temperature distribution in large-chamber refrigerators may be
quite inhomogeneous, the temperature differences reaching up to
10
°C. Similar dispersion can be found inside deep-freezing boxes
(e.g. in those used for blood preservation by deep freezing).
Summary of Contents for U0110
Page 5: ...IE LGR Uxxxx 16 5 General view of a Uxxxx series datalogger U0141 model...
Page 64: ...64 IE LGR Uxxxx 16 Dimensions U0110 U0111 U0122...
Page 65: ...IE LGR Uxxxx 16 65 U0121 U0141...
Page 66: ...66 IE LGR Uxxxx 16 U0246 U0141T U0541 U5841 U6841 U7844 U2422...
Page 67: ...IE LGR Uxxxx 16 67 U3121 U3120 U4130...
Page 68: ...68 IE LGR Uxxxx 16 U3430 U4440 U3631...
Page 69: ...IE LGR Uxxxx 16 69 U8410...