U0246
_________________________________________
Measured values:
3 x input for temperature measurement using
thermocouples
1 x temperature from an external COMET Pt1000/0
probe
1 x internal temperature
Range and input levels:
Inputs 1 to 3 can be switched by the user to measure
temperature using thermocouples or to measure voltage
(-70 to +70) mV.
Parameters of inputs 1 to 3 for different settings:
•
Type „K“ thermocouple (NiCr-Ni)
Range: (-200 to
1300) °C
Accuracy (without probes):
± (|0.3 % of the measurement value|+ 1.5 °C)
Display resolution: 0.
1 °C
Cold junction: compensated in the temperature
range of
(−30 to 70) °C
•
Type „J“ thermocouple (Fe-Co)
Range: (-200 to
750) °C
Accuracy (without probes):
± (|0.3 % of the measurement value| + 1.5 °C)
Display resolution: 0.
1 °C
Cold junction: compensated in the temperature
range of
(−30 to 70) °C
•
Type „S“ thermocouple (Pt10%Rh-Pt)
Range: (0 to
1700) °C
Accuracy (without probes):
± (|0,5 % of the measurement value| + 5 °C)
Display resolution: 0.
1 °C
Cold junction: compensated in the temperature
range of
(−30 to 70) °C
•
Type „B“ thermocouple (Pt30%Rh-Pt)
Range: (250 to
1800) °C
Accuracy (without probes):
± (|0.5 % of the measurement value| + 5 °C)
Display resolution: 0.
1 °C
Cold junction is not compensated.
•
Type „T“ thermocouple (Cu-CuNi)
Range: (-200 to
400) °C
Accuracy (without probes):
± (|0.3 % of the measurement value| + 1.5 °C)
Display resolution: 0.
1 °C
Summary of Contents for U0110
Page 5: ...IE LGR Uxxxx 16 5 General view of a Uxxxx series datalogger U0141 model...
Page 64: ...64 IE LGR Uxxxx 16 Dimensions U0110 U0111 U0122...
Page 65: ...IE LGR Uxxxx 16 65 U0121 U0141...
Page 66: ...66 IE LGR Uxxxx 16 U0246 U0141T U0541 U5841 U6841 U7844 U2422...
Page 67: ...IE LGR Uxxxx 16 67 U3121 U3120 U4130...
Page 68: ...68 IE LGR Uxxxx 16 U3430 U4440 U3631...
Page 69: ...IE LGR Uxxxx 16 69 U8410...