56
IE-LGR-Uxxxx-16
Response time (air flow 1 m/s approx.)
(*2)
:
Internal temperature: t
63
< 2 min, t
90
< 8 min
(temperature step
20 °C)
Relative humidity: t
63
< 45 s, t
90
< 4 min, t
99
<4 h.
(Measured at constant temperature for 30%
humidity step)
External
temperature
input:
Is
determined
by
the response time of the used probe (it valid if
Measuring interval
is set to 1 s).
Display resolution:
Internal temperature incl. dew-
point temperature: 0.1 °C
Relative humidity: 0.1 %RH
External temp. input: 0.1
°C, 16-bit converter range
Recommended calibration interval:
1 year
(*2)
Important information is provided in the chapter
“Models produced – U3631”
U4130
_________________________________________
Measured values:
Internal temperature, relative humidity and barometric
pressure. Dew-point temperature calculated from
the internal temperature and relative humidity.
Ranges:
Temperature: (-30 to +7
0) °C
Relative humidity: (0 to 100) %RH without permanent
condensation
(*2)
.
Barometric pressure (absolute): 600 hPa to 1100 hPa
Dew-point temperature: (-60 to +70)
°C
Accuracy:
Temperature: ± 0.4 °C
Relative humidity:
-
sensor accuracy ±1.8 %RH at 23 ºC in the
range of (0 to 90) %RH
-
hysteresis < ±1 %RH
- non-
linearity < ±1 %RH
- temperature error: see diagrams in
Barometric pressure: ± 1.3 hPa at 23 °C
Dew-
point temperature: ±1.5 °C at an ambient
temperature of T< 25
°C and RH >30 %.
For details see diagrams in
Summary of Contents for U0110
Page 5: ...IE LGR Uxxxx 16 5 General view of a Uxxxx series datalogger U0141 model...
Page 64: ...64 IE LGR Uxxxx 16 Dimensions U0110 U0111 U0122...
Page 65: ...IE LGR Uxxxx 16 65 U0121 U0141...
Page 66: ...66 IE LGR Uxxxx 16 U0246 U0141T U0541 U5841 U6841 U7844 U2422...
Page 67: ...IE LGR Uxxxx 16 67 U3121 U3120 U4130...
Page 68: ...68 IE LGR Uxxxx 16 U3430 U4440 U3631...
Page 69: ...IE LGR Uxxxx 16 69 U8410...