IE-LGR-Uxxxx-16
71
Error
Description and debugging
Error 20
The source value of the calculated quantities is not available.
Check in the SW whether measurement values are available
(temperature and relative humidity) for this calculated quantity
in question (e.g. the dew-point temperature).
Error 21
Calculation failure, incorrect device calibration.
Have the device repaired.
Error 22
Counter overflow in the case that in the device configuration it is
required to report this as a failure.
Reset the counter by means of the SW, or as the case may be change
the setting for the overflow limit.
Error 23
Thermocouple cold junction temperature measurement error.
Check if the internal temperature measurement channel is turned on for
measurement.
Error 50,
Error 52,
Error 55
Device configuration failure, corrupted configuration.
Read the configuration into the SW, check for the correct setting of all
items, save the configuration in the device again.
Error 51
The Digi/E series probe contains invalid calibration constants.
Get the probe repaired or contact the producer.
Error 56
The value is not defined, the measurement on the given channel is
switched off.
Error 57
The value has not been measured yet, the first measurement is waiting
to be completed.
Error 128
Error 255
Error massages generated during communication with the device.
Repeat the communication, contact the producer.
Annex 2: Connection of the Pt1000/E series probe
connector
Summary of Contents for U0110
Page 5: ...IE LGR Uxxxx 16 5 General view of a Uxxxx series datalogger U0141 model...
Page 64: ...64 IE LGR Uxxxx 16 Dimensions U0110 U0111 U0122...
Page 65: ...IE LGR Uxxxx 16 65 U0121 U0141...
Page 66: ...66 IE LGR Uxxxx 16 U0246 U0141T U0541 U5841 U6841 U7844 U2422...
Page 67: ...IE LGR Uxxxx 16 67 U3121 U3120 U4130...
Page 68: ...68 IE LGR Uxxxx 16 U3430 U4440 U3631...
Page 69: ...IE LGR Uxxxx 16 69 U8410...