IE-LGR-UxxxxM-22
37
•
Periodically in preset interval. The interval is selectable from 5
minutes to once per week and you can set different interval when
battery powered and when it have external power available.
•
Asynchronously (immediately) when preset alarm occur.
For battery
mode you can select „Buffered mode“ option which collects
data into internal buffer and then sends them together in preset interval.
It sends more data at once and saves battery life. On the other hand,
the data is received with higher delay.
In the URL field, enter the destination address where you want to send
messages. For use with COMET Database, type the address
of the SOAP server, see chapter 8.2 Automatic data sending to the
database in Comet Database manual (e.g.
).
To send data to the COMET Cloud storage system, select this item as
the destination server and the URL will be set automatically.
To register, use the COMET Cloud Registration Card supplied with
the device (the activation procedure is described on the card). This will
allow trial (usually 3 months) free access to the COMET Cloud system.
You can then extend this trial period by purchasing a Top-Up Voucher
for COMET Cloud (code SLU503).
If the user wants to realize his own collection system, the manufacturer
can offer php scripts for message processing.
Security
________________________________________
When this service is activated in the device configuration, the user is
prompted to set the password for the highest user -
Administrator
.
Subsequently, the device allows users to activate
User1
,
User2
and
PowerUser
and set their own login passwords.
The rights of each user are listed in the following table:
User
Rights
Administrator
- Edit all parameters of device configuration
- Download and erase recorded data
-
Go to service mode
PowerUser
- Edit device configuration without security
parameters, change device time setting
-
Download and erase recorded data
User2
- Download and erase recorded data
-
Switch On/Off device
User1
- Download recorded data
-
Switch On device only (cannot switch it Off)
From the moment the configuration is saved to the instrument,
the instrument will require a username and password each time it is
connected to the SW. You will not be able to connect the device without
entering this information.
Summary
_______________________________________
Printable summary of all device setting.
Summary of Contents for U0110M
Page 5: ...IE LGR UxxxxM 22 5 Datalogger UxxxxM U0141M model...
Page 60: ...60 IE LGR UxxxxM 22 Dimensions U0110M U8410M U0121M...
Page 61: ...IE LGR UxxxxM 22 61 U0141M U3120M U4440M...
Page 62: ...62 IE LGR UxxxxM 22 U3121M U3631M...
Page 63: ...IE LGR UxxxxM 22 63 U0141TM U0843M U5841M U6841M U7841M...