28
IE-LGR-UxxxxM-22
Setting up the device
After the datalogger has been switched on, it starts measuring, data
recording and evaluating alarms according to the device’s setting up
performed by the user. Complete device setup can be performed by the aid
of the
COMET Vision
user software (hereinafter SW). During configuration
editing the device can work normally, nevertheless the access to some
functions (record download, contemporary configuration editing by other
users) is limited.
How to set up the device by means of a program
•
Connect the datalogger to a computer, run the user SW.
•
After starting the program you will see all currently connected devices
in the device list. Note: if you had SW on first and connected
the datalogger then it will be automatically added to the instrument list.
•
Select the device you want to set up. The Device Home panel displays its
current status and device information.
•
Click on the
Configuration
button. The device configuration will be
downloaded; you will be able to view it.
•
If you change the setup of some item, the SW will pass over
into the editing mode. The other
users’ access to the device will be limited
during the editing procedure.
•
In case of a longer idle time the configuration editing mode will close
automatically.
•
Finally save the new configuration into the device (
Apply changes
).
The device setup from a program (Configuration)
General - Information
______________________________
In this panel basic device information is available. Here is above all its
name (
Device name
), which serves to identify the device
and the measurement values it provides. You can name the device e.g.
according to its location or utilization, or you can leave its original
setting. The maximum length of this description, in which you can use
letters, numerals, underscores and further signs, is 31 characters.
More information, which is indicated in this panel, is the device’s serial
number, model, firmware version and information, whether the model
contains an internal modem to send SMS messages.
Summary of Contents for U0110M
Page 5: ...IE LGR UxxxxM 22 5 Datalogger UxxxxM U0141M model...
Page 60: ...60 IE LGR UxxxxM 22 Dimensions U0110M U8410M U0121M...
Page 61: ...IE LGR UxxxxM 22 61 U0141M U3120M U4440M...
Page 62: ...62 IE LGR UxxxxM 22 U3121M U3631M...
Page 63: ...IE LGR UxxxxM 22 63 U0141TM U0843M U5841M U6841M U7841M...