24
Maintenance
Condair TE
6.5
Removing the Drain Pan and Humidification Module
定
固
解
除
固
定
Figure 8:
Removing the drain pan
To remove the drain pan and humidification module:
1. Remove the humidifier cover. Refer to chapter
2. Remove the drain cap from the drain pan, and remove the white connector of the water leakage
sensor. Refer to chapter
3. Use one hand to support the drain pan (and attached humidification module) from below.
Note: the humidification module, when dry, weighs 5.5 lb (2.5 kg). The weight will be heavier if wet.
a. While supported, loosen the knurled screw on the media stopper with a flat-head screwdriver
(or similar tool).
b. Slide the knurled screw to the “release side”.
c.
Heavy!
Slide the drain pan and humidification module out of the humidifier.
The drain pan and humidification module may now be cleaned (refer to chapter
The electrical components in the humidifier may now be inspected (refer to chapter
The internal plumbing components may now be cleaned (refer to chapter
4. Replace the equipment by following these instructions in reverse order.
Media
stopper
Knurled screw
Left side - Release
Drain pan
Media guide
Support
Humid. module
Drain pan