When selecting CanOpen as communication type, the following speeds are available:
20 Kbit/s
50 Kbit/s
100 Kbit/s
125 Kbit/s
250 Kbit/s
500 Kbit/s
1 Mbit/s
③
Interrogator Slave Address
There you can define the network address of the Interrogator
④
Write Values Into Interrogator
By clicking on this button you confirm the changes in this communication section. Without confirming, all changes will be
dismissed.
⑤
Current communication
Displays the current communication settings of the interrogator.
5.3.8.2
Calibration
This chapter of the software it is for authorized personnel only (installation and maintenance). It allows trained personnel to calibrate the
system using one of the following methods:
With sensor characteristics, Twistlock diameter
With a parameter file, or directly with calibration data (these must be prepared in advance)
With a load test on each Twistlock individually
Verify / adjust the calibration with a known weight container
CAUTION!
This part of the software is for authorized persons only!
If the calibration is not done correctly, the system will not work properly.
Conductix-Wampfler will not be responsible for system malfunctioning if the calibration is not
correctly done.