Operating Instructions
LASSTEC
Weighing System 0521
BAL0521-0001-EN
www.conductix.com
translated document
page 39 of 123
⑦
Calibrate the Twistlock.
Once all the previous points have been edited, you can calibrate by clicking on
Calibrate the Twistlock
. The software then
determines the overall Twistlock coefficient based on the entered values and stores it in the interrogator. The calculated
coefficient is then displayed as
Twistlock coefficient
.
5.3.8.2.2
Advanced Calibration
Here you can calibrate the Twistlock either by importing a calibration parameter file or by entering the coefficients for each Twistlock
manually. This calibration method affects each Twistlock individually.
ADVICE!
This method is mainly used for testing purpose as the calibration parameter file as well
as the coefficients of each Twistlock have to be generated in advance with an already
calibrated Lasstec system which used the same Twistlock as the ones which have to
be calibrated now.
①
Kp/Kt coefficient – Current Value.
Displays the current value of the Kp/Kt coefficient stored in the interrogator.
②
Kp/Kt coefficient – Default Value
Displays the factory value of the Kp/Kt coefficient stored in the interrogator.