20
WARNING: Failure to follow the maintenance schedule above could result in reduced instru-
ment performance or overheating of the handpiece or attachment. In addition, heavy use of the
handpiece exceeding the recommended duty cycle can cause the handpiece to overheat. Refer to
“4.0 TECHNICAL SPECIFICATIONS and ACCESSORIES” on page 29 for the associated
handpiece duty cycle. Overheating can lead to possible burn injury to the patient or medical
personnel. Rotation of handpiece usage per day will assist with proper performance.
Table 2: Maintenance Schedule
Catalog (Months)
Number
Product Description
6
12
6020-026
MicroPower High Speed Drill (Removable Lever)
•
6021-026
MicroPower High Speed Drill (Permanent Lever)
•
1375-032
20° Angle Attachment
•
1375-033
Extra-Long 20° Angle Attachment
•
1375-034
70° Contra Angle Attachment w/ Dental Head
•
1375-035
70° Contra Angle Attachment w/ Surgical Head
•
1375-036
90° Angle Attachment
•
Guards
1375-012
MicroPower Medium Bur Guard
•
1375-011
MicroPower Long Bur Guard
•
1375-023
MicroPower Cloward Extra-Long Bur Guard
•
1375-020
MicroPower Laminectomy Guard
•
1387-022
MicroPower Tissue Retractor Guard
•