65
Table 3:
Guidance and manufacturer’s declaration – electromagnetic immunity
The Infrared Thermometer is intended for use in the electromagnetic environment specified
below. The customer the user of the Infrared Thermometer should assure that it is used in such
environment.
Immunity test
IEC 60601 test level
Compliance level
Conducted RF
IEC61000-4-6
3 V
0,15 MHz – 80 MHz
6 V in ISM bands between
0,15 MHz and 80 MHz
3 V
0,15 MHz – 80 MHz
6 V in ISM bands between
0,15 MHz and 80 MHz
Radiated RF
IEC61000-4-3
3 V/m 80 MHz- 2.7 GHz
3 V/m80 MHz- 2.7 GHz
NOTE 1
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2
These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be considered.
If the measured field strength in the location in which the Infrared Thermometer is used
exceeds the applicable RF compliance level above, the Infrared Thermometer should be
observed to verify normal operation. If abnormal performance is observed, additional measures
may be necessary, such as reorienting or relocating the Infrared Thermometer.
Table 4
:
Guidance and manufacturer’s declaration - electromagnetic Immunity
The [Code SI] is intended for use in the electromagnetic environment specified below. The customer or the
user of the [Code SI] should assure that it is used in such an environment
Radiated
RF
IEC61000
-4-3
(Test
specificati
ons
for
ENCLOS
URE
PORT
IMMUNI
TY to
Test
Frequency
(MHz)
Band a)
(MHz)
Service a) Modulation b)
Modulation
b) (W)
Distance
(m)
IMMUNITY
TEST LEVEL
(V/m)
385
380
–390
TETRA
400
Pulse
modulation b)
18 Hz
1,8
0,3
27
450
380
–390
GMRS
460,
FRS 460
FM c)
±
5
kHz
deviation
1 kHz sine
2
0,3
28
710
704
–
787
LTE Band
13,
Pulse
modulation b)
0,2
0,3
9
745