30
In print setup interface, click “Adv-opr” to enter advanced setup interface:
Type
Item
Options
Description
General
setup
Print
Mode
[Auto 4×3]/[Auto
3×4+1]/[Auto 3×4]/[Auto
2×6+1]/[Auto 2×6]/[Auto
3-2+1]/[Auto 3-2][Rhythm
4]/[Rhythm 3][/Rhythm
2]/[Manual]
The system takes the selected option
as default print mode.
Lead
Gain
[Smart]/[Current]
The option selected will be used as
printing gain mode. "Smart" means
the system will adjust gain
automatically to fit paper height;
"Current" means it will use screen
waveform gain as that of printing.
Auto
Strip
[3 sec]/[4 sec]/[5 sec]/[6
sec]/[8 sec]/[10 sec]/[15
sec]/[20 sec]/[25 sec]/[30
sec]
The system takes the selected option
as the print time length of each strip.
Rhythm
Strip
[10 sec]/[15 sec]/[20 sec]/[25
sec]/[30 sec]
When “Print Mode” is set to
“Rhythm 2” “Rhythm 3” or “Rhythm
4”, the system takes the selected
option as the print time length of
each waveform.
Average
QRS
[2×6]/[2×6+
Mark]/[3×4]/[3×4+
Mark]/[4×3]/[4×3+
Mark]/[None]
When “Print Mode” is set to “Auto”
or “Rhythm”, the system uses the
selected format to print the average
QRS waveform.
Auto-Di
agnosis
[All]/[Data]/[Conclusion]/[N
one]
The diagnosis contains data and
conclusion two parts, which can be
chosen by user as demand.
Period
[per1Min]/[per
During the ECG acquisition process,