31
2Min]/[per3Min]/[per5Min]/
[per10Min]/[per20Min]/[per
30Min]/[per 60 Min]/[Off]
the system will automatically
activate the printing operation
according to the selected time
interval. When the printing mode is
manual mode, the printing will
output “Auto 3×4+1” format,
otherwise, it will output according to
the current setting mode.
Advanc
e setup
Print-Sa
ve
[Print and Save]/[Print no
Save]/[Save no Print]
User can choose to print or save the
case after pressing PRINT button
during sampling.
Note 1: The settings of auto strip, rhythm strip, average QRS, auto diagnosis and
periodic print are only optional in auto mode and rhythm mode.
Note 2: If print time length is less than 8s, the sampling and analysis time will be 8s; if
the print time length is equal to or greater than 8s, the sampling and analysis time keeps
the same with print time.
7.12 Lead placement
In the main interface, click
to view the schematic diagram of lead placement, the
European Standard is shown as below:
Click any button to exit.
7.13 About
In the main interface, click
to view the information about the device, which includes
name of device, software version, name of our company, copyright and contact information.