52
12-lead or 15-lead is to evaluate the performance of the automatic ECG analyzer. In addition to
the normal data, the database also includes clinically confirmed ECGs of variety cases, such as
left ventricular hypertrophy, right ventricular hypertrophy, every part of myocardial infarction
and ventricular hypertrophy accompanying myocardial infarction. The database has made a great
contribution to the study of electrocardiology, which is, the CSE group published a report on the
recommended standard for general ECG measurements based on the investigation and study of
the database, which has been widely recognized by the world.
CSE database diagnostic items:
Item
Number
Normal
382
Left ventricular hypertrophy
183
Right ventricular hypertrophy
55
Biventricular hypertrophy
53
Anterior myocardial infarction
170
Inferior myocardial infarction
273
Complex myocardial infraction
104
Synthetical accuracy
1220
4.4 Customized data
4.4.1 Data description
Customized
data
Description
Total recording
number
549
Race
Yellow race
Coverage of
age, gender
Aged from 17 to 87, average age 57.23, standard deviation 21.32;
326 male, average age 55.54, standard deviation 19.81;
223 female, average age 59.70, standard deviation 22.63.
Sampling data
12-lead ECG data (I,
Ⅱ
,
Ⅲ
, AVR, AVL, AVF, V1, V2, V3, V4, V5, V6),
sampling frequency of each channel: 1kHz, amplitude quantization:
2.4μV/LSB.
Remark
The interpretation conclusion of customized data is determined by the
physician diagnostic results of cardiac catheterization and ultrasonic
examination, and the ECG judgment result in physical examination, the
details as blow:
1) Normal ECG
Determined by the diagnostic result that judged as normal in cardiac
catheterization and ultrasonic examination, and the result that judged as