54
Note:
The heart abnormalities such as posterior myocardial ischemia, early posterior MI and old
posterior MI are not included in the database. These abnormalities and other heart disorders not
contained in above sheet won’t be regarded as the judgment object for the verification of
automated interpretation accuracy.
4.6 Data preprocessing
4.6.1 CTS preprocessing
The 16 cases (CAL05000, CAL10000, CAL15000, CAL20000, CAL20002, CAL20100,
CAL20110, CAL20160, CAL20200, CAL20210, CAL20260, CAL20500, CAL30000,
ANE20000, ANE20001, ANE20002) from CTS-ECG shall be processed for voltage conversion
and frequency conversion for resampling as the applicable format in the system. Then cases will
be imported to the device. After that, the verification of automated measurement parameters will
be carried on.
4.6.2 CSE preprocessing
The cases (MA_0001~MA0125, D_0001~D_1220) from the CSE shall be processed for
voltage conversion and frequency conversion for resampling as the applicable format in the