Appendix II EMC Guidance and Manufacturer Declaration
Table 1:
Guidance and manufacturer’s declaration –electromagnetic emission
The Infrared Thermometer is intended for use in the electromagnetic environment specified
below. The purchaser or the user of the device should assure that it is used in such
environment.
Emission test
Compliance
RF emissions CISPR 11
Group 1
RF emissions CISPR 11
Class A
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/flicker
emissions
IEC 61000-3-3
Not applicable
Table 2:
Guidance and manufacturer’s declaration-electromagnetic immunity
The Infrared Thermometer is intended for use in the electromagnetic environment specified
below. The purchaser or the user of the Infrared Thermometer should assure that it is used in
such environment.
Immunity test
IEC60601 test level
Compliance level
Electrostatic discharge (ESD)
IEC 61000-4-2
±8kV contact
± 15 kV air
±8kV contact
±15kV air
Electrical fast transient/burst
IEC 61000-4-4
±2kV for power supply
lines
± 1 kV for input/output line
±2kV for power supply lines
Not Applicable
Surge
IEC 61000-4-5
±1 kV lines to lines
±2 kV lines to earth
±1 kV lines to lines
±2 kV lines to earth
Voltage dips, short interruptions
and voltage vatiations on power
supply input lines
IEC 61000-4-11
<5%UT(>95%dip in UT)
for 0.5 cycle
40% UT(60%dip in UT)
for 5 cycle
70%UT(30%dip in UT) for
25 cycle
<5%UT(>95%dip in UT)
for 5 sec
<5%UT(>95%dip in UT) for
0.5 cycle
40% UT(60%dip in UT) for
5 cycle
70%UT(30%dip in UT) for
25 cycle
<5%UT(>95%dip in UT) for
5 sec
Power frequency (50 / 60Hz)
magnetic field
IEC 61000-4-8
30 A/m
30A/m