52
53
TECHNICAL INFORMATION
C
ONTOUR
®
N
EXT
ONE blood glucose monitoring system
NOTE: If you put the new batteries in the meter within 5 minutes
of taking the old batteries out, the meter saves all your settings
and results. Always check the date and time after you replace
the batteries. If they are correct, other settings and results are
also saved.
4.
Make sure the ‘+’ sign is facing
down on the new batteries.
5.
Hold the battery holder so that the
round end is on the right.
6.
Insert the edge of one battery into
the left side of the compartment
and press down on the right side.
7.
Press the other battery into the left compartment.
8.
Slide the battery holder back into place.
9.
Discard batteries according to your local environmental regulations.
WARNING
Keep batteries away from children. Many types of batteries
are poisonous. If swallowed, immediately contact your
poison control center.
Error Detection Displays
The meter screen displays error codes (
E
plus a number) for test result
errors, strip errors, or system errors. When an error occurs, the meter
beeps 2 times and displays an error code. Press the
OK button
to
turn off the meter.
If you experience continued errors, contact Diabetes Care Customer
Service. See the back cover for contact information.
Error Code
What it Means
What to Do
Strip Errors
E 1
Too Little Blood
Remove the strip. Repeat
the test with a new strip.
E 2
Used Test Strip
Remove the strip. Repeat
the test with a new strip.
E 3
Strip Upside Down
Remove the strip and insert
it correctly.
E 4
Wrong Strip Inserted
Remove the strip. Repeat the
test with a C
ONTOUR
N
EXT
test strip.
E 6
Moisture Damaged Strip
Remove the strip. Repeat
the test with a new strip.
E 8
Strip or Test Errors
Repeat the test with a new
strip. If the error persists,
contact Customer Service.
Testing Errors
E20
Testing Error
Repeat the test with a new
strip. If the error persists,
contact Customer Service.
E24
Too Cold to Test Control
Solution
Move the meter, strip, and
control solution to a warmer
area. Test in 20 minutes.
85096877_CntrNxtONE_UG_EN.indd 52-53
9/21/16 3:47 AM