2) Press                 to confirm and enter the bump menu

1) While the unit is off press & hold                                     to enter the user mode

BUMP TEST

Equipment Needed:

0.25 LPM regulator
Non-absorbent tubing

Calibration adaptor
Calibration gas

If prompted on the unit start-up, follow the on-screen display instructions

3) Confirm the gas concentration values match the gas 

cylinder
4) Attach the calibration adaptor and apply gas

Refer to the manual for more information

+

Note the correct orientation of the calibration adaptor

5) Press                 to start the 30 second countdown

UNIT MUST BE TURNED ON FOR AT LEAST 45 SECONDS BEFORE

PERFORMING A BUMP TEST

If  any  on  the  sensors  fail  the  bump  test,  the  unit  will  begin  an  automatic
calibration and a further 60 second countdown will apply

6) Sensors will show 'P' for pass on a successful bump test

7) Press                 button to display 'start' then press 

                               again to turn the unit on