Kit Installation
2-4 February 2016 | 256574-ENG
R01
2.2.2 TC Kit Installation
Open the chamber door, remove the kit package, and inspect the chamber and kit parts. Contact
Conviron immediately is any damage is found, or if any parts are missing.
2.2.2.1 Installing the Back Wall Plenum Bottom Support
This support is very important since it closes off the bottom of the chamber plenum and directs the
airflow to the four air shelves at equal pressure. If this support is installed upside down, the airflow
will not reach the bottom air shelf. If the support is installed too loosely, the airflow will not be
evenly distributed between the four shelves.
To install the back wall plenum bottom support:
Install the support, Item A, onto the back wall of the chamber with four (4) screws, Item C (Figure
2-2) and tighten the screws to full resistance.
Figure 2-2
TC Back Wall Plenum Bottom Support Installed
The back wall plenum bottom support and the back wall plenum may be installed at
the factory
Install the support with the large flange pointing upward as shown in Figure 2-2.
C
A