2.3.2.1 Installing the Canopy Filter and Fans
The PG kit uses ambient air to cool the light canopy. Air is drawn through the filter on the left side
of the chamber and through the light canopy by the fans on the right side of the chamber. The air
absorbs the heat generated by the canopy lamps and it is then expelled back to atmosphere.
This circulation enables the maximum light output throughout the entire temperature range, since
the lamps are cooled by ambient air, and eases the heat load on the refrigeration system.
To Install the canopy filter:
1. Remove the air inlet cover located in
the left side wall and the insulated
fans covers located in the right side
wall (Figure 2-9). Ensure these three
cavities are empty and clean.
2. Install the air filter into the air inlet
using two (2) screws, Item C.
a. Ensure the filter mesh sits flush
with the inside wall.
b. Tighten the screws to full
resistance (Figure 2-10).
Figure 2-9
PG Fan and Air Inlet Cavities
Figure 2-10
PC Canopy Air Inlet Filter Installed
Chambers ordered for the North American markets will come with the canopy fans
pre-installed by Conviron.
Fan Cavities
Air Inlet Cavity