Alarm Function
Display Alarm Codes
SCl - Temperature probe Pl is short circuit
S01 - Temperature probe Pl is open circuit
SC2 - Temperature probe P2 is short circuit error
S02 - Temperature probe P2 is open circuit error
HPl - Temperature probe Pl is too high
LPl - Temperature probe Pl is too low
HP2 - Temperature probe P2 is too high
LP2 - Temperature probe Pl is too high
do - The door has been left open over 3 minutes
High / Low Temperature Alarm
If the temperature recorded by Pl is higher than the preset value ( factory set at
8°C) for more than 10 minutes, an alarm will sound and the display will flash
HPl
If the temperature recorded by P2 is higher than the preset value ( factory set at
8°C) for more than 10 minutes, an alarm will sound and the display will flash
HP2
The cause of the rise in temperature should be investigated immediately.
Usually the cause is the door being left open or not being shut fully. This can
also happen if the fridge has just been restocked with goods at a higher
temperature than the fridge. It is recommended to check the fridge to
ensure that the fridge is returning to the normal temperature after the alarm
has been activated. If you want to turn the alarm off, P[y}
button.