Low Temperature Alarm
If the temperature recorded by P1 is lower than the preset value ( factory set at
-1 °C) for more than 10 minutes, an alarm will sound and the display will flash
HP1
If the temperature recorded by P2 is lower than the preset value ( factory set at
-1 °C) for more than 10 minutes, an alarm will sound and the display will flash
LP2
The cause of the drop in temperature should be investigated immediately.
Usually the cause is that extremely cold goods have been placed in the fridge.
The fridge should be monitored to ensure it is returning back to the correct
temperature.
Remote Alarm
In the event of a low, high temperature or door alarm, the alarm can be
connected remotely to the red alarm connector at the back. The 4 pin
connector is not to be used as this is factory use only.
Note: the maximum rating for this connection is 1A 30Vdc