28
Immunity Test
IEC 60601
Test Level
Compliance
Level
Electromagnetic
Environmental Guidance [Notes 1 & 2]
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 kHz to 80 MHz
3 V/m
80 MHz to 2 .5 GHz
3 V
3 V/m
Portable and mobile RF communications
equipment should be used no closer to any
part of the CooperSurgical LEEP PRECISION
Integrated System, including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency
of the transmitter .
Recommended separation distance
d
=
d
=
d
=
where
P
is the maximum output power rating
of the transmitter in watts (W) according to
the transmitter manufacturer and
d
is the
recommended separation distance in
meters (m) .
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
a
,
should be less than the compliance level in
each frequency range
b
.
Interference may occur in the vicinity of
equipment marked with the following symbol:
[ ]
3 .5
V
1
√
P
[ ]
3 .5
E
1
√
P
80 MHz to 800 MHz
[ ]
7
E
1
√
P
800 MHz to 2 .5 GHz
NOTE 1:
At 80 MHz and 800 MHz, the higher frequency range applies .
NOTE 2:
These guidelines may not apply in all situations . Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people .
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcasts and TV broadcasts cannot be
predicted theoretically with accuracy . To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered . If the measured field strength in
the location in which the CooperSurgical LEEP PRECISION Integrated System is used exceeds the
applicable RF compliance level above, the CooperSurgical LEEP PRECISION Integrated System
should be observed to verify normal operation . If abnormal performance is observed, additional
measures may be necessary, such as reorienting or relocating the CooperSurgical LEEP PRECISION
Integrated System .
b
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m .
Summary of Contents for LEEP PRECISION LP-20-120
Page 24: ...24 CUT mode 16 1 Power Output Characteristics ...
Page 25: ...25 CUT mode BLEND mode COAG mode 16 2 Power Output at Various Load Resistance ...
Page 52: ...52 16 1 Eigenschaften der Ausgangsleistung ...
Page 53: ...53 16 2 Leistung bei variablem Lastwiderstand ...
Page 80: ...80 16 1 Características de la potencia de salida ...
Page 81: ...81 16 2 Potencia de salida con diferentes resistencias de carga ...
Page 108: ...108 16 1 Caractéristiques de sortie de puissance ...
Page 109: ...109 16 2 Sortie de puissance à résistance de charge variable ...
Page 136: ...136 16 1 Caratteristiche della potenza in uscita ...
Page 137: ...137 16 2 Potenza in uscita a diverse resistenze di carico ...
Page 164: ...164 16 1 Características de potência de saída ...
Page 165: ...165 16 2 Potência de saída a várias resistências de carga ...
Page 192: ...192 16 1 功率輸出特徵 ...
Page 193: ...193 16 2 各種負載電阻下的功率輸出 ...
Page 199: ...Notes Notizen Notas Notes Note Notas 備註 198 ...