9
WARNINGS
• THIS DEVICE IS A BF-TYPE DEVICE. BF-TYPE EQUIPMENT IS B-TYPE DEVICE WITH AN F-TYPE
APPLIED PART.
A B-Type device is piece of equipment providing a particular degree of protection against electric shock,
particularly regarding:
• Allowable leakage current
• Reliability of the protective earth connection .
• An F-Type isolated applied part . The applied part is isolated from all other parts of the equipment to
such a degree that the patient leakage current allowable in single fault condition is not exceeded
when a voltage equal to 1 .1 times the highest rated mains voltage is applied between the applied
part and earth .
• AVOIDING PATIENT BURNS
1 . Do not activate the electrode until all components are in place .
2 . Be certain to have an unobstructed path and view .
3 . Use only non-conductive ancillary instruments .
4 . Remove patient jewelry, metal drape clips, (etc .) and needles .
5 . A Patient Return Electrode must be used . The entire area of the Patient Return Electrode should
be reliability attached to the patient’s body and as close to the operating field as possible .
6 . Place Foot Switch in safe position, out of traffic .
7 . Turn console OFF or place in WAIT state when not in use .
8 . Do not use unit on small appendages due to current density . Bipolar techniques may be desirable
for use on small appendages .
• THERMAL INJURY AND DEFECTS OF TISSUE TREATED WITH LOOP ELECTRODES
Possible injury to cervical tissue may include:
1 . Thermal coagulation injury of the cervix, up to one-third the thickness of the normal cervical
epithelium,
2 . Fragmentation of squamous cervical epithelium attributable to long exposure periods along the
excision site that allows heat to dissipate laterally, and
3 . Partial coagulation of the endocervical epithelium because of lateral heat radiation . Loop
Electrosurgical Excision Procedures may also produce thermal defects at the periphery of the
excised tissue and may make histopathologic interpretation difficult or impossible, preventing
accurate diagnosis and evaluation of the need for further treatment .
• FLAMMABLE GASES/SOLUTIONS
1 . When using a flammable preparatory solution, such as alcohol, be certain that the solution has
completely evaporated before using electrosurgery .
2 . Electrosurgery should not be used in the presence of flammable or explosive gases .
3 . It is recommended that only non-flammable agents be used for cleaning and disinfection .
SMOKE FROM PROCEDURES CONTAINS LIVE CELLS THAT MAY PRESENT BIOHAZARDS
(HIV, ETC .); USE A SMOKE EVACUATION DEVICE .
CONSULT THE PACEMAKER MANUFACTURER’S LITERATURE AND THE PATIENT’S PHYSICIAN
BEFORE USING THIS INSTRUMENT ON A PATIENT USING A PACEMAKER .
Summary of Contents for LEEP PRECISION LP-20-120
Page 24: ...24 CUT mode 16 1 Power Output Characteristics ...
Page 25: ...25 CUT mode BLEND mode COAG mode 16 2 Power Output at Various Load Resistance ...
Page 52: ...52 16 1 Eigenschaften der Ausgangsleistung ...
Page 53: ...53 16 2 Leistung bei variablem Lastwiderstand ...
Page 80: ...80 16 1 Características de la potencia de salida ...
Page 81: ...81 16 2 Potencia de salida con diferentes resistencias de carga ...
Page 108: ...108 16 1 Caractéristiques de sortie de puissance ...
Page 109: ...109 16 2 Sortie de puissance à résistance de charge variable ...
Page 136: ...136 16 1 Caratteristiche della potenza in uscita ...
Page 137: ...137 16 2 Potenza in uscita a diverse resistenze di carico ...
Page 164: ...164 16 1 Características de potência de saída ...
Page 165: ...165 16 2 Potência de saída a várias resistências de carga ...
Page 192: ...192 16 1 功率輸出特徵 ...
Page 193: ...193 16 2 各種負載電阻下的功率輸出 ...
Page 199: ...Notes Notizen Notas Notes Note Notas 備註 198 ...