IONIZER USER MANUAL
17
8) High Voltage Power Failure Alarm on/off set
Press FUN and NEXT (10
times
). HVP.A will display. Off is factory default set as
disable alarm mode. Press UP/DOWN to adjust set from off to on. Press FUN
to store. High voltage power failure alarm mode has been On status now.
*It is recommend to cleaning emitter point at every 3 month cycle or longer.
6) Voltage Adjustment
Press FUN and NEXT(6
times
). PoS.V will display.
7.0 kV is factory default set value. Press UP/DOWN or type numeric value up
to 12.0 kV seconds maximum. Press FUN to store this set value.
Press FUN and NEXT (7
times
). NEG.V will display.
-7.0 kV sec is factory default set value. Press UP/DOWN or type numeric
value up to -12.0 kV seconds maximum. Press FUN to store this set value.
7) Cleaning Cycle Alarm Setting
Press FUN and NEXT (8
times
). Cln.A will display. Off is factory default set as
disable cleaning cycle indication mode. Press UP/DOWN to adjust cleaning
cycle option from off to on.
Press NEXT, then user can adjust cleaning frequency from 1 month to 5
months option. Press UP/DOWN to adjust value and press FUN to store.
7
Setup and Operation