background image

P16.XY20 Piezo Nanopositioning Scanner User Manual

12

7.1 Problem Handling

FAQ are shown in the table below.

7. Maintenance and Treatment

Problem description

 Possible reason

Solution

No displacement or 

less displacement

The connection wire is 

not connected properly or 

poor contact

Check the connection wire

Excessive load 

Load according to the 

product’s load capacity

Sensor zero drift

Check the sensor

Low accuracy

Mountingsurface unevenness

Mount the platform on a flat 

surface

Loose connection

Fasten connection

Wrong connection

Check if the wiring is correct

Vibration or inaccurate 

positioning at startup

Incorrect input control 

parameters

Stop immediately and check the 

parameter Settings

Resonance occurs

Make drive frequency less than  

resonance frequency

Table 2 FAQ and solutions

Attention:

Please handle with care when opening the package, so as not to knock against the 

product and cause permanent damage to it.

Please do not use heavy tools or rough way to open.

Please do not discard all the packaging materials which are needed for repair and 

return.

Summary of Contents for Piezo Nanopositioning P16.XY20

Page 1: ...P16 XY20 Piezo Nanopositioning Scanner User Manual Version V1 0 This document describes the following products P16 XY20S P16 XY20K...

Page 2: ...l If you disassemble or transform the product without permission our company will not be responsible for any consequences arising therefrom To change the product model and other false sales of our com...

Page 3: ...Motion direction 5 4 Operating environment 6 5 Installation and operation 6 5 1 Installation and Precautions 6 5 2 Load 7 5 3 Operating 10 6 Transportation and inspection 11 6 1 Transportation 11 6 2...

Page 4: ...and meaning DANGER Improper handling may could cause injury to the operator Attention Improper handling can cause damage to the equipment No order between items 1 3 Users of This Manual The default u...

Page 5: ...sually if you need a longer cable please contact CoreMorrow A change in temperature and pressure will charge the piezoelectric actuator and the actuator will remain energized for a period of time when...

Page 6: ...nge mechanism has advantages of no friction high positioning accuracy Flexible hinge is an element not affected by static and dynamic friction based on the elastic deformation of metal material e g st...

Page 7: ...echnology interferometry quality detection optical detection nanometer positioning microscopy and imaging image processing and other fields with a good development prospect Y FIG 1 Motion Direction of...

Page 8: ...y check should be carried out before installation such as whether the connection wire is damaged or broken whether the drive power supply is grounded etc 5 1 Installation and Precautions Installation...

Page 9: ...onnector with the corresponding controller according to the identification The wiring mode of LEMO connector is shown as below Safety washer Flat washer Platform Cable connector Ground protection Scre...

Page 10: ...on of the product we do not recommend that the load is too high If a high load must be used please appropriately low the position of the center of gravity of the load The wrong mounting way is shown i...

Page 11: ...expose the product and scratch the product affecting the motion accuracy of the product Short bolts will lead to insufficient compactness of the connection and easy looseness which will damage the int...

Page 12: ...vant controller operations can be carried out 1 Please read the user manual of the controller carefully before operation and then start the controller 2 Set the controller output voltage to zero first...

Page 13: ...shall not exceed the voltage 150V range of P16 XY20 If any abnormal noise or oscillation occurs during use turn off the power immediately and check the parameter setting 6 2 Open Box Inspection Inspec...

Page 14: ...e unevenness Mount the platform on a flat surface Loose connection Fasten connection Wrong connection Check if the wiring is correct Vibration or inaccurate positioning at startup Incorrect input cont...

Page 15: ...product from the controller before cleaning the surface of the product When cleaning gently wipe the surface of the product with a towel dipped in a small amount of detergent and then clean it with a...

Page 16: ...ector The model is LEMO ERA 00 250 CTL as shown below The sensor connector adopts the 4 contacts lemo model LEMO ERA 0s 304 CLL as showne below Note Open loop products have only PZT drive line no sens...

Page 17: ...ontroller model Software driver version Computer operating system that installs supporting software Harbin Core Tomorrow Science Technology Co Ltd Tel 86 451 86268790 Fax 86 451 86267847 Email info co...

Reviews: