background image

3

In order to prevent damage to the product caused by improper operation, the 

following points should be paid attention to when in use.

P16.XY20 piezo nanopositioning scanner is used in a dust-free, oil-free and lubricant-

free environment.

As the product is designed with flexible structure, it is recommended that the load 

should not exceed load capacity of product. When loading, DO NOT apply torque 

force and DO NOT pry or twist the load surface, so as not to damage the structure. 

Recommended voltage range is 0~120V for long-term use and lifetime.

Do not disassemble the product to avoid damage.

Avoid stretching and bending the cable interface to prevent damage to the cable.

Use the dedicated cable provided by CoreMorrow to connect P16.XY20 piezo 

nanopositioning scanner to controller.

Do not connect via an extension cable casually, if you need a longer cable, please 

contact CoreMorrow.

A change in temperature and pressure will charge the piezoelectric actuator, and 

the actuator will remain energized for a period of time when it is disconnected 

from the controller,So the operator should pay attention to when in use.

Do not disassemble P16.XY20 piezo nanopositioning scanner.

Discharge P16.XY20 before installation, which can be done by connecting to the 

controller.

Do not unplug the wire connecting with controller during operation.

Before use, check whether the connection wire of the product is in good condition and 

whether the product and the controller are effectively grounded. During operation, 

strictly standardize the operation. Do not touch the product (up to 150V voltage) with 

your hand when power on, in case of danger.

2. Safety Guide

P16.XY20 piezo nanopositioning scanner is designed with the most advanced technology 

and safety standards for your own safety and the correct use of the product, please 

observe the following points when in use.

Summary of Contents for Piezo Nanopositioning P16.XY20

Page 1: ...P16 XY20 Piezo Nanopositioning Scanner User Manual Version V1 0 This document describes the following products P16 XY20S P16 XY20K...

Page 2: ...l If you disassemble or transform the product without permission our company will not be responsible for any consequences arising therefrom To change the product model and other false sales of our com...

Page 3: ...Motion direction 5 4 Operating environment 6 5 Installation and operation 6 5 1 Installation and Precautions 6 5 2 Load 7 5 3 Operating 10 6 Transportation and inspection 11 6 1 Transportation 11 6 2...

Page 4: ...and meaning DANGER Improper handling may could cause injury to the operator Attention Improper handling can cause damage to the equipment No order between items 1 3 Users of This Manual The default u...

Page 5: ...sually if you need a longer cable please contact CoreMorrow A change in temperature and pressure will charge the piezoelectric actuator and the actuator will remain energized for a period of time when...

Page 6: ...nge mechanism has advantages of no friction high positioning accuracy Flexible hinge is an element not affected by static and dynamic friction based on the elastic deformation of metal material e g st...

Page 7: ...echnology interferometry quality detection optical detection nanometer positioning microscopy and imaging image processing and other fields with a good development prospect Y FIG 1 Motion Direction of...

Page 8: ...y check should be carried out before installation such as whether the connection wire is damaged or broken whether the drive power supply is grounded etc 5 1 Installation and Precautions Installation...

Page 9: ...onnector with the corresponding controller according to the identification The wiring mode of LEMO connector is shown as below Safety washer Flat washer Platform Cable connector Ground protection Scre...

Page 10: ...on of the product we do not recommend that the load is too high If a high load must be used please appropriately low the position of the center of gravity of the load The wrong mounting way is shown i...

Page 11: ...expose the product and scratch the product affecting the motion accuracy of the product Short bolts will lead to insufficient compactness of the connection and easy looseness which will damage the int...

Page 12: ...vant controller operations can be carried out 1 Please read the user manual of the controller carefully before operation and then start the controller 2 Set the controller output voltage to zero first...

Page 13: ...shall not exceed the voltage 150V range of P16 XY20 If any abnormal noise or oscillation occurs during use turn off the power immediately and check the parameter setting 6 2 Open Box Inspection Inspec...

Page 14: ...e unevenness Mount the platform on a flat surface Loose connection Fasten connection Wrong connection Check if the wiring is correct Vibration or inaccurate positioning at startup Incorrect input cont...

Page 15: ...product from the controller before cleaning the surface of the product When cleaning gently wipe the surface of the product with a towel dipped in a small amount of detergent and then clean it with a...

Page 16: ...ector The model is LEMO ERA 00 250 CTL as shown below The sensor connector adopts the 4 contacts lemo model LEMO ERA 0s 304 CLL as showne below Note Open loop products have only PZT drive line no sens...

Page 17: ...ontroller model Software driver version Computer operating system that installs supporting software Harbin Core Tomorrow Science Technology Co Ltd Tel 86 451 86268790 Fax 86 451 86267847 Email info co...

Reviews: