X75 Fusion Splicer
Splicing
Operating Instructions
revision: 06/01
4-7
•
The fusion parameters have been set correctly.
•
Core eccentricity relative to the outside contour must be minimal.
Chapter 6 provides examples of poor splices, possible causes and appropriate remedies.
Re-Fusing
Re-fusing can be used to optimize parameters and to enhance splicing results, if the
selected fusion time was too short, or if fusion current was too low.
-
If the
key is activated, “Fuse?” or “Fuse again?” appears at the display.
This provides you with the opportunity to re-fuse the splice, i.e. to re-ignite the arc.
Splice loss is displayed after re-fusing.
Tensile Test
If the tensile test has been activated, the request “Tensile test?” appears at the display
after fusion.
-
Press the
key to start the tensile test. If the tensile test should not be performed
press
key
-
“Remove fibers” appears at the display after testing.
NOTE
If the tensile test has been activated, the splice loss value is not stored to the splice mem-
ory until after completion of the tensile test. If you open the electrode flap without perform-
ing the tensile test, the splice loss value is not stored to memory.
Summary of Contents for X75 8000 Series
Page 8: ......
Page 18: ...Operating Instructions revision 06 01 X75 Fusion Splicer X75 Fusion Splicer Overview 1 8 ...
Page 58: ...Operating Instructions revision 06 01 X75 Fusion Splicer Splicing 4 10 ...
Page 82: ...Operating Instructions revision 06 01 X75 Fusion Splicer Supplements Changes Notices 7 2 ...
Page 83: ...X75 Fusion Splicer Supplements Changes Notices Operating Instructions revision 06 01 7 3 ...
Page 84: ...Operating Instructions revision 06 01 X75 Fusion Splicer Supplements Changes Notices 7 4 ...