Operating Instructions
revision: 06/01
X75 Fusion Splicer
Troubleshooting and Maintenance
6-14
Matchheads
Electrodes contaminated, arc
burns irregularly.
Clean electrodes.
Fusion current is far too high.
Reduce in steps of 0.5 to
1.0 mA.
Pre-fusion time is far too long. Set pre-fusion parameters to
default values from preset pro-
gram.
Pre-fusion current is far too
high.
Set pre-fusion parameters to
default values from preset pro-
gram.
Autofeed too small.
Increase autofeed in steps of
1
µ
m.
Z-gap too large.
Reduce Z-gap in steps of
1
µ
m.
Symptom
Possible Cause
Remedy