CONFIGURING A RDC/ER-210 USING CHECKMATE™ DL
This RDC Already
Con
fi
gured As ID# xx
Cont Exit
If replacing the existing ID with the new con
fi
guration is desired, press
Cont (F1)
(see display below), otherwise
Exit (F1)
to return to the
RDC Con
fi
guration
display to begin con
fi
guration of a different ID:
RDC # xx
Con
fi
gured
Exit
Press
Exit (F4)
to return to the
Con
fi
guration Options
display.
For Four (4) Channel RDC units, each probe con
fi
guration (channel) must be
individually con
fi
gured. Follow the procedure above to con
fi
gure the
fi
rst channel.
At the completion of the above procedure, the Checkmate DL should be at the
Con
fi
guration Options
display. Now, press
RDC (F2)
to proceed to the
RDC
Con
fi
guration
display and repeat the procedure to con
fi
gure the remaining
channels. Also refer to the
Con
fi
guration of 4 Channel RDC
section below. The
minimum read time interval for a Four (4) Channel RDC is 15 minutes. Although
Corrosometer probes attached to RDC – COT may be set up to have a 5 minute
read time interval, 4 Channel RDC will change the read time to 15 minutes (RDC
– COT (one channel) will retain the 5 minute read interval).
NOTE: RDC units will not collect data while the Checkmate DL is
connected. Make sure to unplug the Checkmate DL from the RDC
to begin data collection cycle.
Updating RDC/ER-210 Con
fi
guration upon Probe Replacement
The initial
Check
reading can be reset easily on the RDC/ER-210 when a
probe is replaced. The
Check
reading is a measure of probe functionality or
integrity. The initial value for CORROSOMETER probes is 800 ±50 divisions. It
is recommended that a
Check
reading be taken and recorded immediately after
unpacking a probe as it will be the value to which all subsequent
Check
readings
will be compared. The general rule is that the
Check
reading should not vary by
more than 1% (±10 divisions) from the initial value. A change greater than 1%
indicates a compromise in the probe integrity and replacement is required.
To reset the
Initial Check
reading, connect the Checkmate DL to the RDC/ER-
210 with the provided LEMO connector cable and power up the Checkmate DL.
Press
SetUp (F4)
from the
Standby
display and then select
RDC (F2)
from the
Con
fi
guration Options
display to proceed to the
RDC Con
fi
guration
display:
Summary of Contents for Checkmate DL
Page 6: ...iv TABLE OF CONTENTS ...
Page 8: ...ii FIGURES AND TABLES ...
Page 10: ...2 INTRODUCTION Figure 1 2 Checkmate DL Instrument Specifications ...
Page 18: ...10 BASICS OF CHECKMATE DL ...
Page 22: ...14 INITIALIZING CHECKMATE DL TO WORK WITH CORRDATA SYSTEM ...
Page 40: ...32 CONFIGURING A RDC ER 210 USING CHECKMATE DL ...
Page 46: ...38 READING A RDC ER 210 WITH THE CHECKMATE DL ...
Page 52: ...44 TRANSFERRING RDC ER 210 DATA TO THE PC ...
Page 60: ...52 READING A DATA LOGGER WITH THE CHECKMATE DL ...
Page 64: ...56 TRANSFERRING DATA LOGGER DATA TO THE PC ...
Page 66: ...58 TROUBLESHOOTING GUIDE ...
Page 67: ...CHECKMATE DL 59 ...
Page 68: ...60 ...
Page 69: ...CHECKMATE DL 61 ...
Page 70: ...62 ...