5.4.1
SYSTEM PARAMETER SETTING
If machine had replace a new computer board, it needs input machine parameter value into new computer
board. Then do the self calculation to save the setting.
5.4.2 press Fn +C to-
gether
Press B+ or B – to adjust the DF value of offset error.
Press A+ to enter next parameter setting.
Do not change the parameter value, unless the result of balancing is off.
Please contact your reseller or manufacturing for after service before make any change.
5.4.1 turn on the machine and wait this homepage.
Display will shows below
picture
5.4.3 press A
-, A+, Fn key in sequence to enter system parameter setting
Press B+ or B – to adjust the I value of axial error.
Press A+ to enter next parameter setting.
Press B+ or B – to adjust the S value of angle error.
Press A+ to enter next parameter setting.
Press B+ or B – to adjust the RES value of Residual value of showing
Press A+ to enter next parameter setting.
Press B+ or B – to adjust equip the width measuring arm.
Press A+ to enter next parameter setting.
Press B+ or B – to adjust diameter measurement. (1 unit = 0.1 inch).
Press A+ to enter next parameter setting.
Press B+ or B – to adjust laser pointer on or off.
Press A+ to return homepage.
Summary of Contents for SP711N
Page 2: ......
Page 3: ...2 Key combinationoverview ...
Page 4: ......
Page 5: ......
Page 6: ......
Page 7: ......
Page 8: ......
Page 9: ......
Page 10: ......
Page 11: ......
Page 12: ......
Page 13: ......
Page 14: ......
Page 15: ......
Page 16: ......
Page 17: ......
Page 18: ......
Page 19: ......
Page 20: ......
Page 21: ......
Page 22: ......
Page 23: ......
Page 25: ......