Section 9
9.2.2 Thermal RF Settings
The Electrode Setup settings (Section 7.1) determine the number and configuration of electrodes when the
generator delivers Thermal RF output. When Thermal RF output is enabled by the user, it is delivered to all
electrodes selected as “active” on the Electrode Setup screen. The following settings determine the
Thermal RF output characteristics.
•
Mode
:
•
Standard
: Time and Temperature control for the Thermal RF output is determined by the Auto
Ramp, Set Temp, and Set Time settings.
•
Stepped
: Time and Temperature control for Thermal RF output is determined by the Start Temp,
Step Temp Increment, Final Temp, Step Time, and Final Time settings. RF Voltage is controlled
automatically by the generator in order that the Temperatures of active electrodes vary in a
stepwise manner.
•
Stagger Start
: When using multiple electrodes, electrodes will turn on sequentially from low to high
index, spaced in time by this setting value.
•
0 seconds
: All electrodes are energized simultaneous when the user enables Thermal RF
output using the front-panel Start button.
•
5-30 seconds
: After the user enables Thermal RF output using the front-panel Start button, each
electrode is not energized until Stagger Start seconds after the previous, lower-indexed
electrode is energized.
The following settings apply only when Thermal RF Mode = Standard.
•
Auto Ramp
: This setting cannot be set by the user in the One Touch interface. It is implicitly set to
On at all times, except when Automatic Temperature Control is disabled (Set Temp = Off), in which
case it is set to Off and the Thermal RF output Voltage must be controlled manually by the user with
the Output Control Knob.
•
Set Temperature (“Set Temp”):
•
Off
: Automatic Temperature Control is disabled. The generator will not control the Temperature
at any electrode or check for Temperature-related errors. Auto Ramp is set to Off and the
Thermal RF output Voltage must be controlled manually by the user with the Output Control
Knob. Use this setting with electrodes that do not monitor temperature, e.g. Cosman CR
Electrode.
•
37-90 °C
: Automatic Temperature Control enabled. Thermal RF output is automatically adjusted
to target the Set Temp value for each electrode.
•
Set Time:
This setting determines the total amount of time that Thermal RF output is delivered to
each electrode. Once this time interval has elapsed for each electrode, electrode output is disabled
automatically.
Summary of Contents for G4
Page 2: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 1 ...
Page 66: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 67: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 68: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 69: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 70: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 71: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 72: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 73: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 74: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 75: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 77: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 78: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 80: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 81: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 82: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 85: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...
Page 86: ...Cosman G4 Radiofrequency Generator Service Manual SPI 11291 Rev A Section 20 ...