CA 8010 Instruction Manual
- 8 -
Calibration the AlcoConnect 8010
Switch on the device
Press briefly the power button to switch on the CA 8010. This is confirmed by an
audible signal and the display shows the current time.
Calibration mode
Aktivate the calibration mode by pressing the central lateral keys between 6 and 7
+ OK-button (10) at the same time. The display shows „
0 0 0 0
“, and the first two
digits start flashing. Replace the two flashing
00
with the number
20
by pressing
the „+“ key and confirm with
OK.
Now the third and fourth
00
flashes. Replace
these two
00
with
the number
10
by pressing the “+” key.
!
Please
do not
press the OK button or any
other key. After a few seconds the device switches automatically into the calibration
mode.
Settings
Factory pre-set for germany (21) in ‰ and alert 0,50 ‰
Country code:
You can change the country code by pressing the „
+
“ button.
Number 20
Switzerland
Number 21
Germany
Number 23
England
Measuring unit
:
You can change the measuring unit by pressing the „OK“ button..
%
‰
mg/l
Alert level:
You can change the alert level by pressing the „-„ button.
0,20‰
0,30‰
0,50‰
0,80‰