background image

− 3 −

Introduction

 Safe Operation

In order to use the Gas Detector safely, be sure to observe the following items.

*Special Instruction and Warning :

Use Panasonic LR6XJK, Toshiba LR6, or Duracell MN1500 batteries.  Use of any other 

manufacuturer or type will violate intrinsic safety requirement.  Remove and replace the batteries 

only in areas known to be non-hazardous.  

1. Introduction (continued from previous page)

DANGER

  If the gas alarm turns ON, immediately take all necessary measures to 

prevent Oxygen deficiency.

WARNING

  Be sure to turn ON the Gas Detector in clean air. Zero calibration (21vol% 

calibration in the case of oxygen) is done automatically. Therefore, the 

wrong gas concentration will be displayed if zero calibration is done in a 

gas atmosphere.

  Do not close the suction port or exhaust port, or no gas detection will be 

possible.

  Use the filter element in a clean state. Normal detection cannot be performed 

if the filter element is dirty or moisture adheres to the filter element.

CAUTION

  Remove the batteries and store the Gas Detector if it is not to be used for a 

long time.

  The  Gas  Detector  has  an  explosion-proof  construction.    Do  not 

disassemble or modify the Gas Detector or change the structure or electric 

circuits of the Gas Detector.  Otherwise, the explosion-proof performance 

of the Gas Detector will be damaged.

  Do not leave the Gas Detector in high-temperature or highly humid places, 

or the performance of the Gas Detector may be adversely affected.

  Keep the Gas Detector away from radical temperature or humidity 

changes, walkie-talkie, mobile phone, or its performance may be adversely 

affected.

  Keep the Gas Detector away from great atmospheric pressure changes, 

or the performance of the sensor may be adversely affected or the sensor 

may be damaged.

  Do not drop, hit, or apply a strong mechanical shock to the Gas Detector, 

or its performance may be adversely affected.

  The Gas Detector is not drip-proof but the leather case protects the Gas 

Detector from water drips. Keep the Gas Detector away from water drips 

as much as possible.

  Do not let the Gas Detector draw in water.

  Prolonged measurement of combustion exhaust gas will shorten the life of 

the sensor. Do not use the sensor continuously for more than 20 minutes.

  High-density SO2 or Cl2 gas may shorten the life of the sensor or cause 

considerable errors in the sensor.

  Do not use the Gas Detector under a pressure of two atmospheres or in 

airplanes, or the Gas Detector may turn ON the alarm due to the change of 

oxygen partial pressure.

Summary of Contents for XP-3180

Page 1: ...the use of two models listed on the left Please keep this instruction manual available for quick reference when needed Before use carefully read this instruction manual and fully understand the conten...

Page 2: ...arm Buzzer Silence Settings 13 Logging 14 Data Logger Communications Screen 15 Time Settings 16 4 Error Message Malfunction Alarm 17 5 Replacement of Consumable Parts 18 Filter Element Replacement 18...

Page 3: ...Mbytes or more CD ROM drive A CD ROM drive capable of reading CD R data The software is provided in a CD R disc USB port A USB1 1 or higher grade port available to Windows which allows connection wit...

Page 4: ...antistatic clothing and electrically conductive shoes antistatic work shoes The floor should also be electrically conductive with an electrical leakage resistance of 10 M or less In the measurement of...

Page 5: ...or modify the Gas Detector or change the structure or electric circuits of the Gas Detector Otherwise the explosion proof performance of the Gas Detector will be damaged Do not leave the Gas Detector...

Page 6: ...ampling hose connection port Connects to the gas sampling hose 3 Exhaust port Exhausts sampled gas 4 Buzzer opening Sounds a buzzer 5 Gas type seal Displays the model number and the types of detectabl...

Page 7: ...the time of a gas alarm G Displays gas concentration in bar graph form H Indicates that the Gas Detector is logging saving data in memory 10 LCD sub screen Displays gas concentration and messages I In...

Page 8: ...ted 4 Suction attachment pipe AT 3A Gas suction port 1 m gas sampling hose for XP 3180 2 Component Names and Functions continued from previous page Cooling Drain Filter Set for XP 3180E No Name Functi...

Page 9: ...ame Function 1 Shoulder belt mounting bracket Connects to the provided shoulder belt 2 Sampling pipe holder Holds the suction attachment pipe XP 3180 3 Fastener Used to extract the Gas Detector or rep...

Page 10: ...count down while the Gas Detector is warming up The LCD sub screen will display a clock 3 When the sensor is stabilized the buzzer will beep and the gas concentration screen will be displayed WARNING...

Page 11: ...replace the sensor unit If the Gas Detector is not turned on in clean air E may be displayed regardless of the lifetime condition of the sensor Always turn ON the Gas Detector in clean air 21vol Cali...

Page 12: ...h is pressed at the time of a gas alarm If the gas concentration reaches the alarm level again the gas alarm buzzer will sound again If the pointer scales out when detecting high concentration gas a d...

Page 13: ...on reaches the alarm level P 13 3 Detectable gas selection XP 3180 and XP 3180E cannot be set 4 LCD selection XP 3180 and XP 3180E cannot be set 5 Execution and completion of logging Logs saves in mem...

Page 14: ...nation to flashing 3 Press the LIGHT swit ch to change from OFF to ON 4 Press the MENU switch to enter the settings The ON indication will change from flashing to steady illumination and 1PE will chan...

Page 15: ...tion to flashing 3 Press the BZ STOP switch and change from ON to OFF will change to 4 Press the MENU switch to enter the settings The OFF indication will then change from flashing to steady illuminat...

Page 16: ...s the LIGHT switch to change OFF to ON 4 Press the MENU switch to enter the settings The ON indication will then change from flashing to steady illumination and 5 LG will change from steady illuminati...

Page 17: ...op flashing and ON will start flashing Data logger communication with the PC will then be possible Operate the PC for data logger communication 4 When communication is finished press the POWER switch...

Page 18: ...the setting and press the MENU switch to enter the value 5 Date Setting When the MENU switch is pressed CL3 will change from flashing to steady illumination and the date value will change from steady...

Page 19: ...tage drops and the remaining battery level is low The Gas Detector will not work when the battery voltage is low Replace the batteries see pages 19 Use approved batteries Duracell MN1500 Panasonic LR6...

Page 20: ...ment with a new one and reassemble the parts XP 3180E 1 Remove the drain cup of the cooling drain filter 2 Remove the O ring on the end of the drain cup 3 Replace the FE 10 Filter Element with a new o...

Page 21: ...into the case part No 3620584053 and zip it shut before entering the hazardous location WARNING Use Panasonic LR6XJK Toshiba LR6 or Duracell MN1500 batteries Use of any other manufacturer or type wil...

Page 22: ...ARNING Ensure there are no explosive gas atmospheres present before opening the cover or replacing the sensor Ensure the new sensor is the correct part number manufactured by New Cosmos Electric Co Lt...

Page 23: ...the gas sampling hose and check that the motor sound of the pump disappears and that Err P is displayed If Err P is not displayed the sensor packing may not have been attached Open the top cover and...

Page 24: ...1 m long gas sampling hose and replace the filter element with a new one see page 18 Remaining battery level If the remaining battery level is low replace the batteries with new ones see page 19 Airti...

Page 25: ...ries again in the correct polarity B a t t e r y r e p l a c e m e n t P19 Battery life Replace the batteries The buzzer does not sound The instrument has been set to the Silent mode Deactivate the Si...

Page 26: ...non conforming product or refund of the original purchase price of the non conforming product In no event will New Cosmos be liable for any other special incidental or consequential damages or losses...

Page 27: ...Ex d ib IIB T4 Gb Other major alarms Remaining battery level sensor error and flow rate drop Main functions 21vol calibration The automatic 21vol calibration of the sensor is possible at the time of...

Page 28: ...Ed Rev 2009 12 08 UL 60079 11 5th Ed Rev 2011 05 05 UL 60079 1 6th Ed Rev 2009 04 10 EN 60079 0 2012 EN 60079 0 2012 EN 60079 11 2012 EN 60079 1 2007 IEC 60079 0 6th Ed Corr 1 IEC 60079 0 6th Ed Corr...

Page 29: ...for the variation in ambient temperature 21 vol adjustment To adjust meter in fresh air to 21 vol indication Explosion proof A structure applied to electrical equipment to prevent the equipment from i...

Page 30: ...MEMO...

Page 31: ......

Page 32: ...If this instruction manual is lost contact New Cosmos Electric or the nearest branch or sales office shown below A copy of the instruction manual will be delivered to you A fee is charged XP 3000ATEX...

Reviews: