Cee® Apogee™ Developer Manual
Page
12
of
15
3.6. Start Up
*For detailed software information, please refer to the DataStream™ manual
1.
Install the tool as shown earlier in Section 4.
2.
Press the lighted power switch to turn the tool power on.
3.
The display should momentarily show boot screen and then show main screen.
4.
Enter the username and password.
(1)
The default username and password are “admin” and “admin2” respectively.
5.
The tool will login and display the Process page.
6.
Navigate to Tools- Manual Control to run system checks to ensure the tool is working properly.
(1)
Turn vacuum on. Use your gloved hand to ensure vacuum is coming through the hole
of the shaft.
(2)
Turn vacuum and off and check that the vacuum is turned off at the hole of the shaft.
(3)
Close the lid and set the spin speed to 1000 rpm. Make sure that the shaft spins and
that the tool is reading 1000 rpm.
(4)
Raise the lid to check the lid interlock. The tool should stop spinning.
(5)
Ensure that the temperature/humidity sensor is reading on the parameter list.
(6)
If equipped with optional dispenses, ensure that dispenses are triggering.
(7)
If equipped with an optional programmable exhaust, ensure it is working properly.
7.
Check that the spin chuck(s) with the tool work properly.
(1)
The spin chucks have a slot that mates with the pin of the spin shaft.
(2)
Position the spin chuck so the slot and the pin line up and press onto the shaft by
hand.
(3)
Ensure that the pin fully engages the slot.
(4)
If equipped with a screw, insert screw and tighten.
(5)
Test vacuum of the spin chuck with a substrate using manual control. Set the vacuum
threshold to 64 kPa. If vacuum is less than 64 kPa, the chuck passes inspection.
(6)
Turn vacuum off and remove the substrate.
(7)
Removal of spin chuck is reverse of installation.
8.
If the tool passes these checks, the tool is ready to operate. Begin by creating a recipe.