Covaris
LE220 (-plus, R-plus, and Rsc) Setup & Instruction Manual
11
Part Number: 010398 Rev H | November 2019
3.
Getting Started
3.1
Fill the Water Bath
The instrument’s transducer must be immersed in a water bath to transmit high-frequency, focused
acoustic energy to the sample tube. For best performance, employ only clean, distilled or deionized
water in the bath.
Open the front door, remove the tank and the first time your system is operated, fill the bath initially
about half full. The first time the WCS and chiller is operated, a substantial amount of air is forced
from the WCS into the bath, which may cause a full bath to overflow. This only occurs the first time an
empty WCS is used.
Once the WCS is filled with water, the bath should be filled to a level of 15. If the transducer is in the
raised position, out of the bath, fill to 15 on the “Fill” side of the gauge. If the transducer is submerged,
fill to 15 on the “Run” side of the gauge.
If the AutoFill option is enabled, after the initial filling of the bath, AutoFill will maintain a sufficient water level in the bath
automatically.
3.2
Start the System
Turn the instrument power ON using the power switch located on the rear, right side of the instrument at the
power entry module. When power is ON, the green door-lock release switch will illuminate and the ventilation
fans in the rear panels will run.
Power on the WCS and CH05 chiller. SonoLab software will perform its temperature control; it is not necessary
to adjust temperature set points from the CH05 control panel.
The first time the chiller and WCS are operated, the WCS filter and UV treatment chamber are empty, and this air is purged as described
above. This is normal, and only occurs the first time the chiller/WCS assembly are operated.
Start the system computer. The computer employs a Windows 10 (or later) Operating System.
SonoLab is an application used to control a Covaris Focused-ultrasonicator for processing a biological or chemical sample. It is designed to
create, edit, and run a processing program called a method. It has provisions to run, pause or stop a method and view execution histories
which contain a record of key instrument parameter values during a run. It also contains editors for customizing system settings and
performing maintenance functions.
SonoLab is installed on the computer purchased with the instrument. No further installation or configuration is required.
Find the SonoLab icon on the desktop:
If the icon is not visible, or if SonoLab is not installed, please refer to
Appendix B
or contact the Covaris Technical Support for installation
instructions.