Covaris
LE220 (-plus, R-plus, and Rsc) Setup & Instruction Manual
27
Part Number: 010398 Rev H | November 2019
-
Dithering
: The LE220 motion system provides dithering motion capability in the X axis (fore-&-aft), Y axis (left-to-right), and Z axis (up-
and-down). If a treatment specifies dithering, click on the “Pen-&-Paper” icon on the right end of the Process bar to display the
Dithering Parameters pane
.
Figure 4.4 Dither Parameter Pane
Enter the specified dithering parameters and click “OK”.
-
Temperature Set Point and Range
: Enter the temperature set point for the protocol and the allowable range into the text entry
fields. SonoLab allows a minimum range of plus/minus 3 degrees.
Figure 4.5 Entering Temperature Set Point and Range from lower to upper limits
Enter Temperature Set Point and Range