Warnings
For the Safety of Operating Personnel:
•
Make sure that the equipment is properly grounded.
DO NOT
operate if it’s not properly grounded.
•
The unit is equipped with a power plug appropriate for the destination country.
DO NOT
, under any circumstances, remove the grounding
prong from the power cord.
•
DO NOT
run an acoustic treatment with the sample cover in the OPEN position or without a water bath - the acoustic system will not
work.
•
If there is any indication that the Safety System is not functioning properly,
DO NOT
operate the equipment and contact Covaris
immediately.
•
If the equipment is used in a manner not specified by the manufacturer, the protection provided by the equipment may be impaired.
To Prevent Damage to the Equipment:
•
The instruments are designed to operate in ambient laboratory conditions e.g., 19 to 25 °C (66 to 77 °F).
DO NOT
operate the instrument in a cold room environment. The system is designed to operate with a water bath and re- circulating
heater/chiller apparatus to control sample temperature.
•
NEVER
run a method without water in the bath since this could damage the transducer. The instrument is equipped with a water level
sensor to protect the transducer and degassing pump. The system will not allow the degassing pump to operate or the acoustic
treatment to start unless an adequate volume of water is detected.
•
Distilled or deionized water should be used to fill the water bath.
•
Unless a Water Conditioning System 3.0 (WCS) is actively employed, empty the water bath and wipe it dry
EVERY DAY
with a lint-free
cloth.
DO NOT
leave water in the tank for an extended time as there is no water filtration or water cleaning system within the apparatus
(unless WCS is employed).
•
DO NOT
employ isopropyl alcohol, ammonia-based or abrasive cleaners in the system, as these will damage the acrylic surfaces.
•
Establish a standard of operation and test the equipment periodically, as described in the maintenance section of this manual.
CAUTION:
All WCS 3 components have pressure ratings above Covaris’ specified maximum operating pressure
which have been extensively tested. However, if plant DI water is connected to the WCS, a catastrophic failure
may result in substantial plant water spillage.
It is the user’s responsibility to specify and implement appropriate
containment/alarm schema for their facility.