LE220 (-plus, R-plus, and Rsc) Setup & Instruction Manual
33
Part Number: 010398 Rev H | November 2019
-
All treatment steps will be applied to all selected rows of the plate once and then repeated for the number of times specified in the
Repeat bar. That is, both programmed Treatments will be given to Row A, then both Treatments will be given to
Row B, and so on until the entire plate is treated and then the treatment steps are repeated.
-
When all method parameters and steps have been defined, save changes.
4.3
Scanning Treatments (LE220Rsc Only)
In a scanning treatment, rather than the sample tray indexing to place one column of the sample tray at a time over the transducer, the
sample tray moves continuously from under the sample tray. In this mode, treatment time is a function of the speed at which the sample tray
moves. One or more rows may be treated, but the rows must be adjacent.
4.3.1 Create a scanning treatment:
-
Click the “
add process
” button.
-
Select
Repeat
from the step type list.
-
Enter the desired number of scans. One scan passes the sample tray over the transducer one time.
-
Click the “
add process
” button again.
-
Select Scan from the Step Type list.
-
Enter the desired treatment parameters as for a normal treatment step.