Covaris
LE220 (-plus, R-plus, and Rsc) Setup & Instruction Manual
42
Part Number: 010398 Rev H | November 2019
6.
System Specifications
Treatment System
Bench-top; high intensity acoustic transducer, temperature monitoring device, circulation pump, water bath with safety
enclosure
Treatment Power
500 Watts Peak Incident Power, 250 Watts Average Power
Dimensions
23” W x 30” D x 19” H (59 cm x 76 cm x 48 cm)
Weight
Approximately 110 lbs. (50 kg)
Power Requirements
100-240 VAC 500 VA maximum, 50-60Hz
Ambient Temperature Range
19 to 25°C (66 to 77°F)
Ambient Humidity Range
30 to 70%
Regulatory Labeling
CE, ETL Mark (for Product Safety), WEEE
Safety
Complies with Low Voltage Directive 2006/95/EC. Certified to IEC/EN/ANSI/UL 61010-1:2004 and CAN/CSA C22.2 No. 61010-1:2004, 2nd
Edition “Safety Requirements for Electrical Equipment for Measurement, Control and Laboratory Use, Part 1: General Requirements”
EMC
Complies with Class A Industrial/Scientific/Medical (ISM) equipment under EN 61326-1:2005, EN 61000-3-2:2004 and EN 61000-3-3:1995
for EU EMC Directive 2004/108/EC. Also FCC Part 15 Class A radio emissions requirements for the USA and ICES-003 Class A for Industry
Canada.
Water Bath
Distilled or deionized water only
Temperature Alarms
Can be programmed from +0.5 °C to +55.0 °C
Computer
Notebook computer
Operating System
Microsoft Windows 10
Application Software
Covaris SonoLab
Data Input
Keyboard, mouse