9.0 The COVVI Power Supply
9.8 Fitting the COVVI Power Supply
The Control Panel should only be fitted by a certified prosthetist.
The information below is to be used as a guide, individual prostheses will differ from patient to patient.
We recommend the Control Panel is placed on the upper side of the prosthetic socket away from any area
which may be loadbearing when carrying heavy items. Lamination dummies are provided if recesses are
required in the outer socket wall.
Control Panel Aperture
Apply the cutting template to the top side of the prosthetic socket.
Carefully cut along the red outline with a cutting disc.
File down any sharp edges.