Ref.No.
032006
TMINI12
(with white LED Flashlight)
Thermometer Operating Instructions
For Model with thermocouple socket
The thermometer is a non-contact infrared thermometer. There are many mathematical modes for the Infrared function. Please
remember to keep away from children and don’t use it for safety related applications.
(Default Screen)
Simply aim the hermometer
at the measure target with
Lens (
7
) and press Meas.
key (
6
) to display the surface temperature.
The Distance:Spot is 12:1. Please make sure the target area is within the field of view.
FUNCTION
Press Mode key (
4
) for scrolling more display function as follows.
Here will show the emissivity data. (The default emissivity is 0.95.)
Press Mode key (
4
) , then press Up key (
2
) or Down key (
3
) to set the emmisivity, then press Mode key (
4
)
to confirm it. The emmisivity can be changed from 0.10 (10E) to 1 (100E).
Press Mode key (
4
) for the Maximum (MAX), Minimum (MIN), Different between MAX and MIN (DIF) and
Average (AVG) modes. During the measurement, the special modes reading will be displayed beside the
mode icon.
Press Up key (
2
) or Down key (
3
) key to change the High Alarm (HAL) or Lo Alarm (LAL), then press Meas.
key (
6
) to confirm it. For example: When the reading 26.9
o
C < LAL 27
o
C, the Low icon will flash and you will hear
a beep sound.
Connect the thermocouple with Thermocouple socket (
5
) and put the probe in/on the target, the thermometer will
display the temperature automatically without pressing any button. To see the minimum or maximum data during the
probe measurement, please hold down the Up key (
2
) or Down key (
3
).
After measure high temp, the probe may remain HOT for a while.
Press Up key (
2
) for LOCK mode ON/OFF. The lock mode is particularly useful for continuous
monitoring of temperatures for up to 60 minutes.
In E,MAX, MIN, DIF, AVG
mode:
Press Down key (
3
) for
o
C or
o
F transferred.
and press Up key (
2
) for backlight function ON/OFF.
In all modes: First hold
on the Meas. key (
6
)
and press Down key (
3
) for laser function ON/OFF.
** The thermometer will automatically shut off if left idle for more than 60sec, unless in PRB mode. (In PRB mode, it will shut off if left
idle for more than 12 minutes.)
CAUTION
1.
WHEN DEVICE IS IN USE, DO NOT LOOK DIRECTLY INTO THE LASER BEAM–PERMANENT EYE DAMAGE MAY RESULT.
2.
USE EXTREME CAUTION WHEN OPERATING THE LASER.
3.
NEVER POINT THE DEVICE TOWARDS ANYONE’S EYES.
4.
KEEP OUT OF REACH OF ALL CHILDREN.
STORAGE & CLEANING
The thermometer should be stored at room temperature between –20 to +65
o
C (-4~149°F).
The sensor lens is the most delicate part of the thermometer. The lens should be kept clean at all times; care should be taken when
cleaning the lens using only a soft cloth or cotton swab with water or medical alcohol. Allowing the lens to fully dry before using the
thermometer. Do not submerge any part of the thermometer.
LCD MESSAGES
The thermometer incorporates visual diagnostic messages as follows:
‘Hi’ or ’Lo’ is displayed when the temperature being measured is outside of the settings of
HAL and LAL.
‘Er2’ is displayed when the thermometer is exposed to rapid changes in the ambient
temperature. ‘Er3’ is displayed when the ambient temperature exceeds 0
o
C (32°F) or +50
o
C
(122°F). The thermometer should be allowed plenty of time (minimum 30 minutes) to stabilize
to the working/room temperature.
For all other error messages it is necessary to reset the thermometer. To reset it, turn the
instrument off, remove the battery and wait for a minimum of one minute, reinsert the battery
and turn on. If the error message remains please contact the Service Department for further
assistance.
BATTERIES
The thermometer incorporates visual low battery indication as follows:
‘Battery OK’: measurements
are possible
‘Battery Low’: battery needs to be
replaced, measurements are still possible
‘Battery Exhausted’: measurements are
not possible
When the ‘Low Battery’ icon indicates the battery is low, the battery should be replaced immediately with AAA, 1.5V batteries.
Please note: It is important to turn the instrument off before replacing the battery otherwise the thermometer may malfunction.
Dispose of used battery promptly and keep away from children.
SPECIFICATION
Item
Non-contact Infrared Scan function
Thermocouple Probe Scan function
(K type; probe not included.)
Measurement Range
-60 to +500
o
C (-76 to +932
o
F)
-64 to +1400°C (-83.2 to +1999°F)
Operating Range
0 to +50°C (32 to +122°F)
Accuracy
(Tobj=15~35°C, Tamb=25°C)
±
1.0°C (1.8°F)
Accuracy
(Tobj=-33~500°C, Tamb=23±3°C)
±
2% of reading or 2°C (4°F)
whichever is
greater
+/-1% of reading or 1°C (1.8°F)
whichever is greater
(Test under Tamb=23±6
C)
Resolution
(-9.9~199.9°C)
0.1°C/0.1°F
Response Time
(90%)
1sec
Distance:Spot
12:1(90% energy covered)
Battery Life
Typ.180, min 140 hours continuous use (Alkaline, without Laser and Back Light.)
Dimensions
31.5 x 94.1 x 94.0mm
(6.9
1.54
2.83 inch)
Weight
204.6 grams including batteries (AAA*2pcs)
Note: Under the electromagnetic field of 3V/m from 200 to 600 MHz, the maximum error is 10
o
C (18
o
F).
Caution: The measure range is for thermometer only. User should choose proper probe types for different kinds of
application. Please make sure the target to be measured will not exceed the temperature range of the
probe to avoid permanent damage of the thermocouple probe.
EMC/RFI
: Readings may be affected if the unit is operated within radio frequency electromagnetic field strength of
approximately 3 volts per meter, but the performance of the instrument will not be permanently affected.
CPS PRODUCTS INC.
1010 E 31
st
Street Hialeah, FL 33013 Tel: 305-687-4121; 1-800-277-3808
www.cpsproducts.com
; email: cpssales@cpsproducts.com