Lamp Replacement
83
8.1.1 Replacing the Transparency Lamp
The transparency fluorescent lamp is located inside the scanner top
cover.
Light level is too low
The following message is displayed on the screen.
To exit the message window, perform one of the following:
Click
Abort
; the scanner application quits. Replace the defective
lamp (see the following instructions),
or
Click
Ignore
, and continue with the pre-scan/scan. Replace the
defective lamp when you have completed your work.
Lamp does not light up
The transparency lamps always light up a few seconds after the scanner
is switched on.
To confirm that a lamp needs replacing, switch the system
OFF
and
ON
again.
If the lamp does not light up after 15 seconds, replace the lamp and
try again.
Summary of Contents for EverSmart
Page 1: ...EverSmart Scanners www creo com scanners 399Z50703F Installation Maintenance English...
Page 2: ......
Page 8: ......
Page 12: ......
Page 13: ...1 Chapter 1 Site Preparation General Requirements 2 Safety Instructions 6...
Page 19: ...7 Chapter 2 Unpacking Unpacking Procedure 8 Unlocking Screws Removing Packing 13...
Page 29: ...Powering on the Scanner 17 2 2 2 2 Make sure that the scanner is switched OFF...
Page 40: ......
Page 41: ...29 Chapter 5 SCSI Configuration Configuration Options 30 Setting the SCSI ID 43...
Page 92: ......
Page 109: ...Lamp Replacement 97 4 4 4 4 Pull out the left side of the panel...
Page 110: ...98 Chapter 8 Maintenance 5 5 5 5 Slide out the right side of the panel...
Page 120: ......
Page 128: ......
Page 129: ...117 Appendix A Packing the Scanner Locking the Scanner 118 Packing the Scanner 121...
Page 137: ...125 Appendix B Technical Specifications...
Page 140: ......
Page 145: ...133 Appendix D Application Log Files...
Page 147: ...135 Appendix E Catalog Numbers...
Page 150: ......