At the end of this process, the LED located on the sensor flashes orange;
please remove the
reference
plate S,
then press once again the measure
key.
If the operation it is uncertain it was done correctly (i.e. reference plate placed
wrongly), it is possible to press the
C
key to exit from the procedure, without
any calibration changes.
Figure 26 - Zero point calibration request
Pressing the measure key, the procedure progress column shows up again.
At the end of the second calibration process, the screen will show the
following:
Figure 27 - End of calibration process
After few sec. the first screen shows up (please see figure 19), requesting to
replace the reference plate over the sensor and to proceed again with the
verification.
If the procedure was correctly performed, the verification will have positive
results, and the calibration procedure will end; if the verification fails it is
possible to repeat the procedure once again.