7.4 Measurement in continuous mode
If the option “
Continuous
” is selected from the settings on the measurement
menu (see paragraph 6.1.0), it is possible to save the measured data
continuously, to create a profile of water distribution along the cross direction
of the forming fabric. The supplied plastic ring (Z), can be placed on the
instrument handle, in between the electronic plug cap and the silicone cover,
to help the proper alignment of the instrument.
Selecting the option “
Continuous
” will turn off the option “
FFT
” from the
settings on the FFT menu.
CAUTION!
If the position where the scan is performed is very noisy
(turbulence), it is suggested to set the speed/accuracy filter to 4
or 5 prior to start measurements in continuous mode.
Data acquisition start after the end of the first measure of the series and the
following screen appear:
Figure 43 - Continuous data acquisition screen
Is possible to save up to 7936 measure sample for a total time of
approximately 60 minutes at a rate of 2 samples per second.
Water mass profile saved in the memory can be reviewed by selecting a
measure with profile data and pressing the CENTER key. A screen like the
following will appear: