TROUBLESHOOTING

 

ATTENTION!

  For  any  replacement  of  parts,  contact  yout  local  dealer.  It’s 

not 

recommended 

the replacement by non-qualified people. 

 

Listed below are the most frequent problems and possible solutions, which could be solved 

by the user.

THE VACUUM PUMP DOES NOT SWITCH ON

•Lack of power -------------------------------------------------
•Disarmed circuit breaker ----------------------------------
•The power cord is not plugged in ----------------------

•When taking out the suction instrument out of the 
support , the vacuum pump does not activate -------

•Fuse burn ------------------------------------------------------

POSIBLE CAUSES

SOLUTION

•Verify if there’s a power failure in your area.
•Rearm the circuit breaker.
•Connect the power cord to the outlet properly.

 •Consult

 

 

“How to proceed when service is needed” 

(Page. 13).

•Change the fuse, which is located at the electronic 

circuit,  inside the command box.  To  replace the 
fuse, unplug the equipment from the outlet, push 
the pins of the plastic rivets (Fig. 8, page 9) inside 
the command box, with a slim tip, remove de rivets 
and the cover, locate the fuse and replace it by a 
new  one.  Take  out  the  the  pushed  plastic  pins, 
place the lid again and the plastic rivets. 

Note: 

To 

place the rivets, insert the plastic pins at the center 
of the rivets, about 2 mm, put the rivets inside the 
holes and push the pins until they get alined with 
the  rivets to  lock  it.  See  “

Safety Devices

”  (Item 1, 

page 9).

If the problem persists after the verification of all items, contact your local dealer.

THE VACUUM PUMP SWITCHES ON BUT PRODUCES INSUFICIENT OR DOES NOT 
PRODUCE VACUUM

•Lack of water in the vacuum pump --------------------

•The residue filter is full ----------------------------------------

•The redisue filter cover isn’t properly closed -----------

•Air entrance through the pipes ---------------------------

•Verify if the water valve is closed or if there’s lack of 

water ar your area.

•Clean  the  residue  filter.  See  “

Preventive 

Maintenance

” (Page. 11).

•Remove the residue filter cover, verify the gasket 

fitting and mount it again.

•Verify  if  there’s  weak  connections  or  leaks  at  the 

dental unit vacuum pipes.

POSIBLE CAUSES

SOLUTION

C R I S T Ó F O L I

V A C U U M  P U M P S

I N S T R U C T I O N 

M A N U A L 

Summary of Contents for SuperGap Evolution 2

Page 1: ... better serve its customers To fulfill all the requirements of the applicable standards promote the continuous improvement of its quality and environmental standards prevent pollution reduce its environmental impact and continuously train its staff achieving therefore a sustainable and profitable maximization of its value Rev 2 CRISTÓFOLI S MISSION CRISTÓFOLI QUALITY AND ENVIRONMENTAL POLICY Crist...

Page 2: ...commendations and Warnings Installation Instructions Safety Devices Component Description How To Use the SuperGap Evolution Vacuum Pump How To Identify Your Autoclave Technical Data Preventive Maintenance Troubleshooting Quality Control How To Proceed When Service Is Needed Guidance For Final Disposal Of The Equipment Warranty Terms 03 04 04 04 05 09 09 09 10 10 11 12 13 13 14 15 ...

Page 3: ...ection Terminal Date of Manufacture 3 Keep Dry Recyclable Keep out of sunlight This side up Fragile handle with care LOT SN Serial Number Lot number ISO 13485 13485 DE S G A E M S E T T Ã S I O S C R I S T Ó F O L I V A C U U M P U M P S I N S T R U C T I O N M A N U A L ...

Page 4: ...ues from odontological treatments or surgeries INTENDED USE RECOMMENDATIONS AND WARNINGS Read all instructions of this manual before using your SuperGap Evolution The incorrect use may result in malfunction and or accidents For your safety when performing any kind of maintenance remember to unplug the power cord and use PPE Personal Protective Equipment Do not allow patients especially kids to get...

Page 5: ... voltage device and it leaves the factory adjusted for 220 V For 127 V operation reposition the selector switch placed at the back of the equipment Fig 9 page 9 See How To Identify Your Vacuum Pump Fig 11 page 10 For installation use a grounded outlet 2 Pins Grounding 20 A plugging phase neutral or phase phase at the side pins and grounding at the middle pin Never plug the grounding on neutral Ele...

Page 6: ...ressure must be around 3 to 40 wcm water column meters above 20 wcm it could be possible to hear an acute sound from the flow regulator Connect the hose with a 90 angle to the water inlet at the front of the vacuum pump Fig 8 page 9 and the other side to the water valve The water flow regulation is automatically done by an internal device Fig 8 page 9 SUCTION The pipe that connects the dental unit...

Page 7: ...tall the noise reducer Fig 3 observing the arrow that indicates flow direction Figs 3 and 4 Attention The incorrect installation of the noise reducer backwards will cause damage to the vacuum pump Connect the R L adapter to the drain exit of the dental unit after it connect the hose to the adapter and fasten it with the clamp Note The vacuum pump and the cuspidor may be connected to the same drain...

Page 8: ... most distant from the dental unit vacuum pump SuperGap 2 10 m max 32 mm Model Pipes length Pipes diameter Table 2 Fig 5 Fig 6 Strip Connector Residue Filter Voltage Selector Switch Identification Label Fig 7 45 C R I S T Ó F O L I V A C U U M P U M P S I N S T R U C T I O N M A N U A L ...

Page 9: ...se it is necessary to wait for 2 minutes until the vacuum system is filled with water To turn on the vacuum pump just take the suction instrument out from it s support and to turn it off get the instrument into the support again so the vacuum pump will be automatically deactivated At the end of the workday clean the residue filter Fig 10 See Preventive Maintenance Page 11 HOW TO USE THE SUPERGAP E...

Page 10: ...ification This equipment is manufactured by Cristófoli Biossegurança company whose the Quality Management System is certified by and in accordance with ISO 13485 2003 I Suitable for up to 4 dental units 14 6 kg 13 5 kg 23 2 cm x 23 5 cm x 35 cm 27 cm x 27 cm x 39 cm 127 220V AC single phase 60 Hz 0 16 l min 560 W 3 4 HP 0 56 h kW 300 liters min 3450 620 mm Hg 78 dB 70 0 a 75 70 106 kPa 5 ºC a 40 º...

Page 11: ...and biodegradable neutral detergent after it clean with a cloth moistened with 70 alcohol and 1 peracetic acid Never use solvents or petroleum derivatives Verify the openings at the bottom of the motor if there s dirtiness accumulation on it perform cleaning YEARLY General revision and flow limiter regulation we recommend that this procedure be performed by an authorized technical assistance Unplu...

Page 12: ...m tip remove de rivets and the cover locate the fuse and replace it by a new one Take out the the pushed plastic pins place the lid again and the plastic rivets Note To place the rivets insert the plastic pins at the center of the rivets about 2 mm put the rivets inside the holes and push the pins until they get alined with the rivets to lock it See Safety Devices Item 1 page 9 If the problem pers...

Page 13: ...e date of manufacture found in the identification label located in the back of the equipment Fig 6 page 8 and a description of the problem Always contact your local dealer If you have problems contacting your dealer contact us through our website www cristofoli com HOW TO PROCEED WHEN SERVICE IS NEEDED THE VACUUM PUMP STOPS AND DOES NOT SWITCH ON The motor is overheated and the thermal fuse was ac...

Page 14: ...EQUIPAMENTOS DE BIOSSEGURANÇA LTDA strongly recommends users of their products to seek the best destination when disposing equipment or components taking into account the materials recycling legislation effective in each country We advise you to take your equipment to specialized recycling companies that due to the continuous and fast paced development of new recycling technologies and materials r...

Page 15: ...ng from natural causes such as floods lightning etc Sinister theft or robbery Damage caused by accidents such as dents drops short circuits fire etc Problems due to failure of the power supply and or building wiring Improper use installation of the equipment and or plugging it into an incorrect voltage outlet Removal and or tampering of the serial number shown in the identification label of the pr...

Page 16: ......

Reviews: